首页> 外文期刊>Current Directions in Biomedical Engineering >Influence of different test gases in a non-destructive 100% quality control system for medical devices
【24h】

Influence of different test gases in a non-destructive 100% quality control system for medical devices

机译:100%无损医疗设备质量控制体系中不同测试气体的影响

获取原文
       

摘要

The purpose of this research is to evaluate the detectability of defect membranes in intravenous (IV) infusion filter systems. The device under test (DUT) protects critical ill patients and has a high priority for the risk management of intensive care units. The developed quality control system stands out from other filter integrity methods because no method located on this topic represents such a simple, reliable, fast and non-destructive technique, examined without liquid. The invented method works as a pressure driven test and uses gas to identify defects. Previous studies have demonstrated the capability of detecting various types of errors. In this paper the influence of different test gases on the detectability of smallest defects is presented.
机译:这项研究的目的是评估静脉(IV)输液过滤系统中缺陷膜的可检测性。被测设备(DUT)保护重症患者,并在重症监护病房的风险管理中具有较高的优先级。发达的质量控制系统在其他过滤器完整性方法中脱颖而出,因为在没有液体的情况下检查的任何方法都无法代表这种简单,可靠,快速且无损的技术。本发明的方法用作压力驱动测试,并使用气体来识别缺陷。先前的研究证明了检测各种类型错误的能力。本文提出了不同测试气体对最小缺陷可检测性的影响。