首页> 外文期刊>Current Directions in Biomedical Engineering >Investigation of the pressure gradient of embolic protection devices
【24h】

Investigation of the pressure gradient of embolic protection devices

机译:栓塞保护装置压力梯度的研究

获取原文
       

摘要

To avoid debris coming to the cerebral vessels during carotid artery stenting, embolic protection devices (EPD) are placed in front of the lesion. To evaluate their influence on the antegrade blood flow a test setup with a silicone tube simulating the internal carotid artery is realized. The pressure gradient of five different EPD was measured while particles were brought into the circuit and were caught by the systems. Additionally the microscopic structure of the systems was observed to correlate the morphology and the pressure gradient. The FiberNet device had the lowest pressure gradient. It was the only system that consists of fibers contrary to the systems RX Accunet, Angioguard RX, FilterWireEZ and EmboshieldNAV that contain porous membranes.
机译:为避免在颈动脉支架置入过程中碎片进入脑血管,栓塞保护装置(EPD)被放置在病变的前面。为了评估它们对顺行血流的影响,实现了带有硅胶管模拟颈内动脉的测试装置。在将颗粒带入回路并被系统捕获时,测量了五种不同EPD的压力梯度。另外,观察到系统的微观结构使形态和压力梯度相关。 FiberNet设备的压力梯度最低。它是唯一由纤维组成的系统,与包含多孔膜的RX Accunet,Angioguard RX,FilterWireEZ和EmboshieldNAV系统相反。