首页> 外文期刊>Clinical ophthalmology >Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and?Orbscan
【24h】

Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and?Orbscan

机译:角膜中央厚度的比较:超声测厚术与裂隙灯光学相干断层扫描,镜面显微镜和Orbscan

获取原文
           

摘要

Background: Central corneal thickness (CCT) can be measured by using contact and non-contact methods. Ultrasound pachymetry (US pachymetry) is a contact method for measuring CCT and is perhaps the most commonly used method. However, non-contact methods like scanning slit topography (Orbscan II), slit-lamp optical coherence tomography (SL-OCT), and specular microscopy are also used. Not many studies have correlated the measurement of CCT with all four modalities. The purpose of this study was to compare and correlate the CCT measurements obtained by US pachymetry with SL-OCT, specular microscopy, and Orbscan. Method: This is a prospective, comparative study done in an institutional setting. Thirty-two eyes of 32 subjects with no known ocular disease and best-corrected visual acuity of 20/20 were enrolled. CCT measurements were obtained using SL-OCT, specular microscopy, scanning slit topography (Orbscan), and US pachymetry. Three measurements were made with each instrument by the same operator. Mean, standard deviation, and coefficient of variation were calculated for CCT measurements acquired by the four measurement devices. Bland–Altman plot was constructed to determine the agreements between the CCT measurements obtained by different equipment. Results: The mean CCT was 548.16±48.68 μm by US pachymetry. In comparison, CCT averaged 546.36±44.17 μm by SL-OCT, 557.61±49.92 μm by specular microscopy, and 551.03±48.96 μm by Orbscan for all subjects. Measurements by the various modalities were strongly correlated. Correlations (r2) of CCT, as measured by US pachymetry compared with other modalities, were: SL-OCT (r2=0.98, P<0.0001), specular microscopy (r2=0.98, P<0.0001), and Orbscan (r2=0.96, P<0.0001). All modalities had a linear correlation with US pachymetry measurements. Conclusion: In subjects with healthy corneas, SL-OCT, specular microscopy, and Orbscan (with correction factor) can be used interchangeably with US pachymetry in certain clinical settings. The four modalities showed significant linear correlations with one another.
机译:背景:中央角膜厚度(CCT)可以通过接触和非接触方法进行测量。超声测厚法(US pachymetry)是一种用于测量CCT的接触方法,也许是最常用的方法。但是,也可以使用非接触式方法,例如扫描狭缝形貌(Orbscan II),狭缝灯光学相干断层扫描(SL-OCT)和镜面显微镜。很少有研究将CCT的测量与所有四种方式相关联。这项研究的目的是比较和关联通过美国测厚仪使用SL-OCT,镜面显微镜和Orbscan获得的CCT测量值。方法:这是在机构环境中进行的前瞻性比较研究。入选了32名无已知眼病且矫正视力为20/20的受试者的32只眼睛。使用SL-OCT,镜面显微镜,扫描狭缝形貌(Orbscan)和US测厚法获得CCT测量值。同一操作员对每种仪器进行了三项测量。计算通过四个测量设备获得的CCT测量的平均值,标准偏差和变异系数。构造Bland-Altman图来确定不同设备获得的CCT测量值之间的一致性。结果:通过美国测厚仪测得的平均CCT为548.16±48.68μm。相比之下,所有受试者的SL-OCT的CCT平均值为546.36±44.17μm,镜面显微镜的CCT平均为557.61±49.92μm,而Orbscan平均为551.03±48.96μm。通过各种方式进行的测量高度相关。通过US测厚仪与其他方式进行比较,CCT的相关性(r2)为:SL-OCT(r2 = 0.98,P <0.0001),镜面显微镜检查(r2 = 0.98,P <0.0001)和Orbscan(r2 = 0.96) ,P <0.0001)。所有模态与美国测厚仪的测量值均呈线性相关。结论:在具有健康角膜的受试者中,SL-OCT,镜面显微镜和Orbscan(具有校正因子)可以在某些临床环境中与US测厚仪互换使用。四种模态显示出显着的线性相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号