首页> 外文期刊>Chemistry central journal >Assessment of an in vitro whole cigarette smoke exposure system: The Borgwaldt RM20S 8-syringe smoking machine
【24h】

Assessment of an in vitro whole cigarette smoke exposure system: The Borgwaldt RM20S 8-syringe smoking machine

机译:体外全烟暴露系统评估:Borgwaldt RM20S 8针吸烟机

获取原文

摘要

Background There have been many recent developments of in vitro cigarette smoke systems closely replicating in vivo exposures. The Borgwaldt RM20S smoking machine (RM20S) enables the serial dilution and delivery of cigarette smoke to exposure chambers for in vitro analyses. In this study we have demonstrated reliability and robustness testing of the RM20S in delivering smoke to in vitro cultures using an in-house designed whole smoke exposure chamber. Results The syringe precision and accuracy of smoke dose generated by the RM20S was assessed using a methane gas standard and resulted in a repeatability error of ≤9%. Differential electrical mobility particle spectrometry (DMS) measured smoke particles generated from reference 3R4F cigarettes at points along the RM20S. 53% ± 5.9% of particles by mass reached the chamber, the remainder deposited in the syringe or connecting tubing and ~16% deposited in the chamber. Spectrofluorometric quantification of particle deposition within chambers indicated a positive correlation between smoke concentration and particle deposition. In vitro air-liquid interface (ALI) cultures (H292 lung epithelial cells), exposed to whole smoke (1:60 dilution (smoke:air, equivalent to ~5 μg/cm2)) demonstrated uniform smoke delivery within the chamber. Conclusions These results suggest this smoke exposure system is a reliable and repeatable method of generating and exposing ALI in vitro cultures to cigarette smoke. This system will enable the evaluation of future tobacco products and individual components of cigarette smoke and may be used as an alternative in vitro tool for evaluating other aerosols and gaseous mixtures such as air pollutants, inhaled pharmaceuticals and cosmetics.
机译:背景技术近来有许多近距离发展的可复制体内暴露的体外香烟烟雾系统​​。 Borgwaldt RM20S吸烟机(RM20S)可以对香烟烟雾进行连续稀释并将其输送到暴露室进行体外分析。在这项研究中,我们证明了使用内部设计的全烟暴露室,RM20S在将烟气传给体外培养物中的可靠性和耐用性测试。结果使用甲烷气体标准液评估了RM20S产生的注射器精度和烟雾剂量的准确性,并导致重复性误差≤9%。差示电迁移率粒子光谱(DMS)测量了从参比3R4F香烟在RM20S沿点产生的烟雾颗粒。按质量计,有53%±5.9%的颗粒到达腔室,其余的沉积在注射器或连接管中,约16%的颗粒沉积在腔室内。腔室中颗粒沉积的荧光光谱定量分析表明,烟雾浓度与颗粒沉积之间呈正相关。暴露于全烟气(1:60稀释度(烟气:空气,相当于〜5μg/ cm2))的体外气液界面(ALI)培养物(H292肺上皮细胞)显示出室内均匀的烟雾输送。结论这些结果表明,这种烟雾暴露系统是一种可靠且可重复的方法,可用于将ALI体外培养物生成并暴露于香烟烟雾中。该系统将能够评估未来的烟草产品和香烟中的单个成分,并且可以用作评估其他气溶胶和气体混合物(例如空气污染物,吸入药物和化妆品)的替代性体外工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号