首页> 外文期刊>Cardiology Journal >Comparable vascular response of a new generation sirolimus eluting stents when compared to fluoropolymer everolimus eluting stents in the porcine coronary restenosis model
【24h】

Comparable vascular response of a new generation sirolimus eluting stents when compared to fluoropolymer everolimus eluting stents in the porcine coronary restenosis model

机译:猪冠状动脉再狭窄模型中新一代西罗莫司洗脱支架与含氟聚合物依维莫司洗脱支架相比可比的血管反应

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background: Novel sirolimus eluting stents (SES) have shown non-inferior clinical outcomes when compared to everolimus eluting stents (EES), however only limited preclinical data have been published. Therefore, we evaluate vascular response of a new generation biodegradable polymer SES (BP-SES: Alex Plus, Balton) and fluoropolymer EES (EES: Xience Pro, Abbott) in the porcine coronary restenosis model. Methods: A total of 40 stents were implanted with 120% overstretch in coronaries of 17 domestic swine: 16 BP-SES, 16 EES and 8 bare metal controls (BMS). Following 28 and 90 days, coronary angiography and optical coherence tomography (OCT) was performed, animals sacrificed and stented segments harvested for pathological evaluation. Results: At 28 days neointimal thickness in OCT was lowest in the BP-SES when compared to EES and BMS (0.18 ± 0.1 vs. 0.39 ± 0.1 vs. 0.34 ± 0.2 mm, respectively; p = 0.04). There was no difference in the proportion of malapposed or uncovered struts, although protruding covered struts were more common in BP-SES (14.8 ± 10% vs. 4.1 ± 4% vs. 3.7 ± 6%; p = 0.03). In pathology, the lowest neointimal thickness was confirmed in BP-SES (p Conclusions: New generation, BP-SES show similar vascular healing and biocompatibility profile with marginally higher degree of restenosis inhibition, when compared to fluoropolymer EES in the porcine coronary restenosis model.
机译:背景:与依维莫司洗脱支架(EES)相比,新型西罗莫司洗脱支架(SES)的临床疗效不差,但是临床前数据很少。因此,我们在猪冠状动脉再狭窄模型中评估了新一代可生物降解的聚合物SES(BP-SES:Alex Plus,Balton)和含氟聚合物EES(EES:Xience Pro,Abbott)的血管反应。方法:共向40只支架植入了17只家猪的冠状动脉中120%的过度拉伸:16例BP-SES,16例EES和8例裸金属对照(BMS)。在28天和90天后,进行冠状动脉造影和光学相干断层扫描(OCT),处死动物并收集支架段以进行病理评估。结果:在28天时,与EES和BMS相比,BP-SES中OCT的新内膜厚度最低(分别为0.18±0.1 vs. 0.39±0.1 vs. 0.34±0.2 mm; p = 0.04)。尽管在BP-SES中更常见的是突出的覆盖式支撑(14.8±10%vs. 4.1±4%vs. 3.7±6%; p = 0.03),但支气管缺失或未覆盖的比例没有差异。在病理学中,BP-SES证实了最低的新内膜厚度(p结论:与猪冠状动脉再狭窄模型中的含氟聚合物EES相比,新一代BP-SES表现出相似的血管愈合和生物相容性特征,再狭窄抑制程度略高。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号