首页> 外文期刊>Brazilian Journal of Medical and Biological Research >Bioelectrical impedance spectroscopy for the assessment of body fluid volumes of term neonates
【24h】

Bioelectrical impedance spectroscopy for the assessment of body fluid volumes of term neonates

机译:生物电阻抗光谱法用于评估足月新生儿的体液量

获取原文
获取外文期刊封面目录资料

摘要

The assessment of fluid volume in neonates by a noninvasive, inexpensive, and fast method can contribute significantly to increase the quality of neonatal care. The objective of the present study was to calibrate an acquisition system and software to estimate the bioelectrical impedance parameters obtained by a method of bioelectrical impedance spectroscopy based on step response and to develop specific equations for the neonatal population to determine body fluid compartments. Bioelectric impedance measurements were performed by a laboratory homemade instrument. The volumes were estimated in a clinical study on 30 full-term neonates at four different times during the first month of life. During the first 24 hours of life the total body water, extracellular water and intracellular water were 2.09 ± 0.25, 1.20 ± 0.19, and 0.90 ± 0.25 liters, respectively. By the 48th hour they were 1.87 ± 0.27, 1.08 ± 0.17, and 0.79 ± 0.21 liters, respectively. On the 10th day they were 2.02 ± 0.25, 1.29 ± 0.21, and 0.72 ± 0.14 liters, respectively, and after 1 month they were 2.34 ± 0.27, 1.62 ± 0.20, and 0.72 ± 0.13 liters, respectively. The behavior of the estimated volume was correlated with neonatal body weight changes, leading to a better interpretation of such changes. In conclusion, this study indicates the feasibility of bioelectrical impedance spectroscopy as a method to help fluid administration in intensive care neonatal units, and also contribute to the development of new equations to estimate neonatal body fluid contents.
机译:通过无创,廉价和快速的方法评估新生儿的体液量,可以显着提高新生儿护理质量。本研究的目的是校准采集系统和软件,以估算通过基于阶跃响应的生物电阻抗光谱法获得的生物电阻抗参数,并开发针对新生儿的特定方程式以确定体液区室。生物电阻抗测量是通过实验室自制仪器进行的。在临床研究中,对生命的第一个月中四个不同时间的30名足月新生儿进行了估计。在生命的最初24小时内,全身水,细胞外水和细胞内水分别为2.09±0.25、1.20±0.19和0.90±0.25升。到第48小时,它们分别为1.87±0.27、1.08±0.17和0.79±0.21升。在第10天,它们分别为2.02±0.25、1.29±0.21和0.72±0.14升,而在1个月后,它们分别为2.34±0.27、1.62±0.20和0.72±0.13升。估计体积的行为与新生儿体重变化相关,从而可以更好地解释这种变化。总之,这项研究表明生物电阻抗光谱法作为一种有助于重症监护新生儿室输液的方法的可行性,并且还有助于开发新的方程式来估算新生儿体液含量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号