Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability

机译:由Hemochron签名精英和Abbott i-STAT进行的心脏手术激活凝血时间的即时护理测量:一致性,一致性和临床可靠性

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance, and anticoagulant reversal. However, concerns exist regarding reproducibility of ACT assays and comparability of devices. We evaluated the agreement of ACT assays using four parallel measurements performed on two commonly used devices each (i.e., two Hemochron Signature Elite (Hemochron) and two Abbott i-STAT (i-STAT) devices, respectively). Blood samples from 30 patients undergoing cardiac surgery on CPB were assayed at specified steps (baseline, after heparin administration, after protamine administration) with four parallel measurements (two of each device type) using commercial Kaolin activated assays provided by the respective manufactures. Measurements were compared between identical and different device types using linear regression, Bland-Altman analyses, and calculation of Cohen’s kappa coefficient. Parallel i-STAT ACTs demonstrated a good linear correlation (r?=?0.985). Bias, as determined by Bland-Altman analysis, was low (??3.8?s; 95% limits of agreement (LOA): ??77.8 -70.2?s), and Cohen’s Kappa demonstrated good agreement (kappa?=?0.809). Hemochron derived ACTs demonstrated worse linear correlation (r?=?0.782), larger bias with considerably broader LOA (??13.14?s; 95%LOA:-316.3–290?s), and lesser concordance between parallel assays (kappa?=?0.554). Although demonstrating a fair linear correlation (r?=?0.815), parallel measurements on different ACT-devices showed large bias (?20s; 95% LOA: ??290-250?s) and little concordance (kappa?=?0.368). Overall, disconcordant results according to clinically predefined target values were more frequent with the Hemochron than i-STAT. Furthermore, while discrepancies in ACT between two parallel iSTAT assays showed little or no clinical relevance, deviations from parallel Hemochron assays and iSTAT versus Hemochron measurements revealed marked and sometimes clinically critical deviations. Currently used ACT point-of-care devices cannot be used interchangeably. Furthermore, our data question the reliability of the Hemochron in assessing adequacy of heparin anticoagulation monitoring for CPB.
机译:由于肝素抗凝作用不足和逆转不足会导致心肺分流(CPB)手术期间发生并发症,因此通过即时护理(POC)激活凝血时间(ACT)测量进行肝素抗凝监测对于CPB的启动,维持和抗凝逆转至关重要。然而,存在关于ACT测定的可重复性和装置的可比性的担忧。我们使用分别在两个常用设备上进行的四个平行测量(即分别是两个Hemochron Signature Elite(Hemochron)和两个Abbott i-STAT(i-STAT)设备)来评估ACT分析的一致性。使用各制造商提供的商业高岭土活化测定法,在指定步骤(基线,肝素给药后,鱼精蛋白给药后)中,对指定步骤(基线,肝素给药后,鱼精蛋白给药后)的30名患者的血样进行了四次平行测量(每种设备类型中的两次)。使用线性回归,Bland-Altman分析和计算科恩的kappa系数,比较了相同和不同设备类型之间的测量结果。并行的i-STAT ACT表现出良好的线性相关性(r?=?0.985)。通过Bland-Altman分析确定的偏倚很低(?3.8?s; 95%的协议限度(LOA):? 77.8 -70.2?s),科恩的Kappa表现出良好的一致性(kappa?=?0.809)。 。 Hemochron衍生的ACTs表现出更差的线性相关性(r?=?0.782),更大的偏倚和更大的LOA(?13.14?s; 95%LOA:-316.3–290?s),以及平行测定之间的一致性较低(kappa == 0.554)。尽管证明了良好的线性相关性(r = 0.815),但是在不同的ACT器件上进行的并行测量显示出较大的偏差(?20s; 95%LOA:?290-250?s)和很少的一致性(kappa == 0.368)。 。总体而言,与i-STAT相比,Hemochron的临床预定义目标值不一致的结果更为常见。此外,尽管两次平行iSTAT测定之间的ACT差异显示很少或没有临床相关性,但平行Hemochron测定以及iSTAT与Hemochron测量的偏差显示出明显的,有时甚至是临床上的严重偏差。当前使用的ACT即时护理设备不能互换使用。此外,我们的数据质疑了Hemochron在评估CPB肝素抗凝监测的充分性方面的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号