首页> 外文期刊>BMC Pulmonary Medicine >Effects of equipment and technique on peak flow measurements
【24h】

Effects of equipment and technique on peak flow measurements

机译:设备和技术对峰值流量测量的影响

获取原文
           

摘要

Background Different lung function equipment and different respiratory manoeuvres may produce different Peak Expiratory Flow (PEF) results. Although the PEF is the most common lung function test, there have been few studies of these effects and no previous study has evaluated both factors in a single group of patients. Methods We studied 36 subjects (PEF range 80–570 l/min). All patients recorded PEF measurements using a short rapid expiration following maximal inspiration (PEF technique) or a forced maximal expiration to residual volume (FVC technique). Measurements were made using a Wright's peak flow meter, a turbine spirometer and a Fleisch pneumotachograph spirometer. Results The mean PEF was 8.7% higher when the PEF technique was used (compared with FVC technique, p Conclusion Peak flow measurements are affected by the instruction given and by the device and Peak Flow scale used. Patient management decisions should not be based on PEF measurement made on different instruments.
机译:背景技术不同的肺功能设备和不同的呼吸操作可能会产生不同的峰值呼气流量(PEF)结果。尽管PEF是最常见的肺功能检查,但很少有关于这些作用的研究,并且以前没有研究对单个患者中的两种因素进行评估。方法我们研究了36名受试者(PEF范围为80–570 l / min)。所有患者均使用最大吸气后的短时快速呼气(PEF技术)或强制最大呼气剩余量(FVC技术)记录PEF测量值。使用赖特的峰值流量计,涡轮肺活量计和弗莱施气动力学肺活量计肺活量计进行测量。结果使用PEF技术时,平均PEF高出8.7%(与FVC技术相比,p结论峰值流量测量受所给的说明以及所用设备和峰值流量标度的影响,患者管理决策不应基于PEF在不同的仪器上进行测量。