首页> 外文期刊>Biological research: BR >Testing of a new prototype surgical stapler that automates the rollover sleeve technique for venous anastomoses
【24h】

Testing of a new prototype surgical stapler that automates the rollover sleeve technique for venous anastomoses

机译:测试新型原型手术吻合器,该吻合器可自动完成翻身套技术以实现静脉吻合

获取原文

摘要

The creation of successful vascular anastomoses is of primary importance in many surgical fields. Numerous attempts to automate this process have been made. These techniques have slowly gained acceptance, but their use is still limited. This report details feasibility testing of a new prototype stapler that automates the rollover sleeve technique for venous vascular anastomoses. Male and female mongrel dogs (n=7) (25-32 kg) were used. A segment of the right (n=5) or left (n=2) iliac vein was harvested for interposition grafts after the contra lateral side was transected. In each dog, two end-to-end venous anastomoses at the interposition grafts were performed. The standard anastomosis employed continuous mattress sutures. The experimental anastomosis was performed with a new prototype surgical stapler. The stapled anastomosis was proximal and the sutured was distal. In all experiments, it was possible to perform the experimental anastomosis with the stapler. Complications included two small leaks, one due to misfiring of a single pin in one experimental site. These leaks required suture reinforcement. One dog died of hemorrhage due to a slipped suture at the vein harvest site. One vein had thrombus seen at the sutured site although no technical abnormalities at either of the anastomoses could be found. After two weeks, grafts were inspected grossly and histologically. Healing appeared normal. There was a trend for less inflammatory cells infiltrating stapled sites; however, this was not statistically significant. The experiments demonstrate that this device can automate the rollover sleeve technique for venous anastomoses.
机译:成功的血管吻合术在许多外科领域中至关重要。已经进行了许多尝试来使该过程自动化。这些技术已逐渐被接受,但其使用仍然受到限制。该报告详细介绍了一种新的原型订书机的可行性测试,该订书机可以自动完成翻倒套技术用于静脉血管吻合术。使用雄性和雌性杂种犬(n = 7)(25-32公斤)。对侧切开后,收获骨右段(n = 5)或左段(n = 2)的一部分用于介入移植。在每只狗中,在移植物中进行了两个端到端静脉吻合术。标准的吻合采用连续的床垫缝合线。实验吻合是使用新的原型外科缝合器进行的。吻合钉在近端,缝合在远端。在所有实验中,都可以用订书机进行实验吻合。并发症包括两次小泄漏,一个是由于在一个实验点上单个针的未击穿引起的。这些泄漏需要缝合线加固。一只狗死于出血,原因是在静脉收集部位的缝合线滑落。尽管在任一吻合处均未发现技术异常,但在缝合部位可见一条静脉血栓。两周后,对移植物进行了全面的组织学检查。愈合似乎正常。炎性细胞浸入固定部位的趋势是减少的。但是,这在统计上并不显着。实验表明,该装置可以自动完成静脉吻合的翻身套技术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信