首页> 外文期刊>Biomedical Optics Express >Simultaneous optical coherence tomography and lipofuscin autofluorescence imaging of the retina with a single broadband light source at 480nm
【24h】

Simultaneous optical coherence tomography and lipofuscin autofluorescence imaging of the retina with a single broadband light source at 480nm

机译:使用480nm的单个宽带光源同时进行视网膜的光学相干断层扫描和脂褐素自发荧光成像

获取原文
           

摘要

We accomplished spectral domain optical coherence tomography and auto-fluorescence microscopy for imaging the retina with a single broadband light source centered at 480 nm. This technique is able to provide simultaneous structural imaging and lipofuscin molecular contrast of the retina. Since the two imaging modalities are provided by the same group of photons, their images are intrinsically registered. To test the capabilities of the technique we periodically imaged the retinas of the same rats for four weeks. The images successfully demonstrated lipofuscin accumulation in the retinal pigment epithelium with aging. The experimental results showed that the dual-modal imaging system can be a potentially powerful tool in the study of age-related degenerative retinal diseases.
机译:我们完成了光谱域光学相干断层扫描和自动荧光显微镜,用一个以480 nm为中心的宽带光源对视网膜成像。该技术能够提供同时的视网膜结构成像和脂褐素分子对比。由于两个成像模态由同一组光子提供,因此它们的图像在本质上是配准的。为了测试这项技术的功能,我们定期对同一只大鼠的视网膜进行了四个星期的成像。这些图像成功地证明了脂褐素随着年龄的增长而沉积在视网膜色素上皮中。实验结果表明,双模式成像系统可以成为研究年龄相关性视网膜变性疾病的潜在强大工具。