首页> 外文期刊>Current Medical Imaging Reviews >Post Fusion MRI Artifacting - An In Vitro Study of Defined Intervertebral Test Spacers
【24h】

Post Fusion MRI Artifacting - An In Vitro Study of Defined Intervertebral Test Spacers

机译:融合后MRI伪影-定义的椎间测试垫片的体外研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: Interbody spacers for anterior spine fusion are made of different materials, which can affect the post-fusion MRI scans. Susceptibility artifacts specially for metallic implants can decrease the image quality. This study focused on the influence of determined implant parameters like shape and implant volume in MRI artifacting independent from selected MRI-sequences and implant materials.nnMaterial and methods: In this study we evaluated the post-implantation MRI scans of 3 determined titanium cuboids and 6 cylinders. MRI scans were carried out by using T1 TSE sequences. The total artifact volume (TAV) of all examined implants was calculated for statistical t-test correlation and implant volume (IV)/TAV-relation.nnResults: Considering all examined test implants with an increasing implant size the TAV became significantly larger (p < 0,001) with simultaneous reduction of the implant artifact relation. Regarding a TAV-correlation for cylinders with an implant volume equivalent to the cuboids significantly differences in MRI artifacting could be demonstrated (p < 0,05).nnConclusion: Susceptibility artifacts are directly influenced by parameters which determine the implant size and volume. With regard to a more advantageous IV / TAV-relation for larger titanium implants the artifact rate was more limited to the implant's direct surroundings. Significant differences with regard to the TAV could be determined in favor of a cylindric implant shape. The results demonstrated that the artifact rate of our used test implants is influenced by the material volume as well as implant shape.
机译:目的:用于前路脊柱融合的椎间间隔器由不同的材料制成,这可能会影响融合后的MRI扫描。专门用于金属植入物的磁化伪影会降低图像质量。这项研究的重点是确定的植入物参数(如形状和植入物体积)对MRI人工成像的影响,而与选定的MRI序列和植入物材料无关.nn材料和方法:在这项研究中,我们评估了3个确定的钛长方体和6个钛长方体的植入后MRI扫描气瓶。使用T1 TSE序列进行MRI扫描。计算所有被检查植入物的总假体体积(TAV)以进行统计t检验相关性和植入物体积(IV)/ TAV关系.nn结果:考虑到所有被检查的植入物尺寸增加的测试植入物,TAV显着增大(p < 0,001),同时减少植入物的假象关系。关于植入物体积等于长方体的圆柱体的TAV相关性,可以证明MRI伪像存在显着差异(p <0.05)。结论:敏感性伪像直接受到确定植入物大小和体积的参数的影响。对于较大的钛植入物而言,更有利的IV / TAV关系是,假象率更多地局限于植入物的直接周围环境。对于TAV,可以确定有利于圆柱植入物形状的显着差异。结果表明,我们使用的测试植入物的假象率受材料体积以及植入物形状的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号