Internal Quality Control for Erythrocyte Sedimentation Rate Measured by TEST-1 Analyzer

机译:用TEST-1分析仪测定红细胞沉降速率的内部质量控制

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The TEST 1~(TM) is a fully automated analyzer for measurement of the erythrocyte sedimentation rate. This system employs a particular capillary where blood is moved by a special hydrodynamic system. This original method is not able to measure stabilized samples for quality control, probably because stabilized erythrocytes offer a higher resistance to movement into the capillary, giving a distorted sedimentation curve. We evaluated whether the stability of collected EDTA samples, as declared by the producer company, was sufficient to use samples measured the day before as internal quality control samples. We also evaluated whether different tubes could modify the test results between stored and fresh samples. The difference between ESRs measured in fresh and stored samples are non-relevant after 24h and 48h, using both the tubes considered. The agreement between fresh and stored samples was better than that obtained by comparison with the Westergren method and can be used for the internal quality control procedure.
机译:TEST 1TM是用于测量红细胞沉降速率的全自动分析仪。该系统采用特殊的毛细管,血液通过特殊的流体动力系统移动。这种原始方法无法测量稳定的样品以进行质量控制,这可能是因为稳定的红细胞对进入毛细管的移动具有更高的抵抗力,从而使沉淀曲线失真。我们评估了生产商公司宣布的收集到的EDTA样品的稳定性是否足以将前一天测得的样品用作内部质量控制样品。我们还评估了不同的试管是否可以修改存储样品与新鲜样品之间的测试结果。使用所考虑的两个试管,在新鲜样品和储存样品中测得的ESR之间的差异在24h和48h后无关紧要。新鲜样品和储存样品之间的一致性好于与Westergren方法进行比较所获得的一致性,可用于内部质量控制程序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号