首页> 外文期刊>Biomedical Engineering, IEEE Transactions on >Evaluation of Model-Enhanced Ultrasound-Assisted Interventional Guidance in a Cardiac Phantom
【24h】

Evaluation of Model-Enhanced Ultrasound-Assisted Interventional Guidance in a Cardiac Phantom

机译:心脏幻影模型增强超声辅助介入指导的评估

获取原文
获取原文并翻译 | 示例
           

摘要

Minimizing invasiveness associated with cardiac procedures has led to limited visual access to the target tissues. To address these limitations, we have developed a visualization environment that integrates interventional ultrasound (US) imaging with preoperative anatomical models and virtual representations of the surgical instruments tracked in real time. In this paper, we present a comprehensive evaluation of our model-enhanced US-guidance environment by simulating clinically relevant interventions in vitro . We have demonstrated that model-enhanced US guidance provides a clinically desired targeting accuracy better than 3-mm rms and maintains this level of accuracy even in the case of image-to-patient misalignments that are often encountered in the clinic. These studies emphasize the benefits of integrating real-time imaging with preoperative data to enhance surgical navigation in the absence of direct vision during minimally invasive cardiac interventions.
机译:使与心脏手术相关的侵袭性最小化导致对目标组织的有限视觉进入。为了解决这些限制,我们开发了一种可视化环境,该环境将介入性超声(US)成像与术前解剖模型和实时跟踪的手术器械的虚拟表示相集成。在本文中,我们通过模拟体外临床相关干预措施,对我们的模型增强型美国制导环境进行了全面评估。我们已经证明,模型增强的美国指南提供了优于3毫米rms的临床上所需的瞄准精度,即使在临床上经常遇到的图像对患者未对准的情况下,也能保持这一精度水平。这些研究强调了将实时成像与术前数据相结合的好处,以在微创心脏介入治疗中在没有直视的情况下增强手术导航。