Image-Based Mechanical Analysis of Stent Deformation: Concept and Exemplary Implementation for Aortic Valve Stents

机译:基于图像的支架变形的力学分析:主动脉瓣支架的概念和示例性实施

获取原文
获取原文并翻译 | 示例
           

摘要

An approach for extracting the radial force load on an implanted stent from medical images is proposed. To exemplify the approach, a system is presented which computes a radial force estimation from computer tomography images acquired from patients who underwent transcatheter aortic valve implantation (TAVI). The deformed shape of the implanted valve prosthesis' Nitinol frame is extracted from the images. A set of displacement vectors is computed that parameterizes the observed deformation. An iterative relaxation algorithm is employed to adapt the information extracted from the images to a finite-element model of the stent, and the radial components of the interaction forces between the stent and the tissue are extracted. For the evaluation of the method, tests were run using the clinical data from 21 patients. Stent modeling and extraction of the radial forces were successful in 18 cases. Synthetic test cases were generated, in addition, for assessing the sensitivity to the measurement errors. In a sensitivity analysis, the geometric error of the stent reconstruction was below 0.3 mm, which is below the image resolution. The distribution of the radial forces was qualitatively and quantitatively reasonable. An uncertainty remains in the quantitative evaluation of the radial forces due to the uncertainty in defining a radial direction on the deformed stent. With our approach, the mechanical situation of TAVI stents after the implantation can be studied in vivo, which may help to understand the mechanisms that lead to the complications and improve stent design.
机译:提出了一种从医学图像中提取植入式支架上的径向力载荷的方法。为了举例说明该方法,提出了一种系统,该系统从计算机断层图像中计算径向力估计值,该计算机断层图像是从经导管主动脉瓣膜植入术(TAVI)的患者那里获取的。从图像中提取植入的瓣膜假体的镍钛诺框架的变形形状。计算出一组位移矢量,该矢量将观察到的变形参数化。采用迭代松弛算法使从图像中提取的信息适应支架的有限元模型,并提取支架与组织之间相互作用力的径向分量。为了评估该方法,使用来自21位患者的临床数据进行了测试。支架建模和径向力提取成功18例。此外,还生成了综合测试用例,用于评估对测量误差的敏感性。在敏感性分析中,支架重建的几何误差低于0.3 mm,低于图像分辨率。径向力的分布在质量和数量上都是合理的。由于在变形的支架上限定径向方向的不确定性,在径向力的定量评估中仍然存在不确定性。通过我们的方法,可以在体内研究TAVI支架植入后的机械状况,这可能有助于了解导致并发症的机制并改善支架设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号