首页> 外文期刊>Australasian physical & engineering sciences in medicine >In-vivo dosimetry from total body irradiation patients (2000 - 2006): results and analysis
【24h】

In-vivo dosimetry from total body irradiation patients (2000 - 2006): results and analysis

机译:全身放射患者的体内剂量测定(2000年至2006年):结果与分析

获取原文
获取原文并翻译 | 示例

摘要

In-vivo dosimetry is an important technique to ensure accuracy of delivered dose during total body irradiation (TBI). We present an analysis of semiconductor diode dosimetry, constituting seven years of dosimetry data from eighty-six patients who underwent total body irradiation. For lateral field irradiation, the mean exit dose, averaged over five anatomical sites (head, sternal notch, chest, abdomen and pelvis) and as a percentage of the planned dose was 95.7 % (SD 7.8 %). For AP/PA irradiation, the mean exit dose averaged over five anatomical sites and as a percentage of the planned dose was 95.5 % (SD 9.8 %). We propose a number of possible reasons for these differences, including patient setup variations, movement of the patient and diodes during treatment, imprecise placement of planned bolus material, inaccurate inhomogeneity corrections and modelling by the treatment planning system.
机译:体内剂量测定是一项重要技术,可确保在全身照射(TBI)期间精确地递送剂量。我们目前对半导体二极管剂量学进行了分析,构成了来自八十六名接受全身照射的患者的七年剂量学数据。对于侧场照射,平均出射剂量在五个解剖部位(头部,胸骨切迹,胸部,腹部和骨盆)平均,占计划剂量的百分比为95.7%(标准偏差为7.8%)。对于AP / PA照射,在五个解剖部位的平均平均出射剂量为计划剂量的95.5%(标准差为9.8%)。我们提出了造成这些差异的多种可能原因,包括患者设置的变化,患者和二极管在治疗过程中的运动,计划大剂量材料的不精确放置,不均匀的不正确校正和治疗计划系统的建模。