首页> 外文期刊>Artificial Organs >Cycle Testing of the MagScrew Total Artificial Heart External Battery Pack: Update I
【24h】

Cycle Testing of the MagScrew Total Artificial Heart External Battery Pack: Update I

机译:MagScrew人造心脏总体外电池组的周期测试:更新I

获取原文
获取原文并翻译 | 示例
           

摘要

MagScrew total artificial heart (TAH) external battery pack (EBP) cycle bench testing continued over a period of 18 months using two fresh Wilson Greatbatch lithium ion EBPs during continuous charge and discharge cycles under a simulated TAH system current requirement. The same electronic load developed for our initial testing was used to simulate the MagScrew current waveforms typically observed during nominal operation. The current load profiles for this test were modified from the ones previously described and applied to the EBP under test during a voltage-defined discharge cycle. The test ended when EBP#2 reached end of life at 1450 cycles. At that point, EBP#1 remained healthy with a capacity of 175?min until full discharge. Performance of EBP#2 was still within expected ranges. Performance of EBP#1 exceeded expectations. These differences are probably caused by slight manufacturing changes. More tests will provide additional data to define a statistical distribution to better characterize EBP performance. In conclusion, endurance performance of the EBP remained satisfactory.
机译:在模拟的TAH系统电流要求下,在连续的充放电循环中,使用两个新鲜的Wilson Greatbatch锂离子EBP,在18个月的时间内继续进行了MagScrew人工心脏(TAH)总体外部心脏电池组(EBP)循环台式试验。为我们的初始测试开发的相同电子负载用于模拟在标称运行期间通常观察到的MagScrew电流波形。此测试的当前负载曲线已从先前描述的修改,并在电压定义的放电周期内应用于测试的EBP。当EBP#2在1450个周期达到使用寿命时,测试结束。到那时,EBP#1保持健康,容量为175?min,直到完全放电。 EBP#2的性能仍在预期范围内。 EBP#1的性能超出预期。这些差异可能是由轻微的制造更改引起的。更多测试将提供更多数据,以定义统计分布,从而更好地表征EBP性能。总之,EBP的耐力表现仍然令人满意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号