Technical pitfall while reducing the mismatch between LCP PLT and upper end Tibia in proximal Tibia fractures

机译:技术性陷阱,同时减少了胫骨近端骨折中LCP PLT和上端胫骨之间的不匹配

获取原文
获取原文并翻译 | 示例
           

摘要

Minimal invasive technique along with the widespread use of the new pre contoured locking plates (LCP PLT, LISS PLT) has given rise many alignment related problems in fractures of proximal tibia. A significant implant related problem with the use of pre contoured locking plate in proximal tibia fractures is the hardware irritation of the subcutaneous tissue by upper end of plate. This is caused by the large profile of the implant which sits off the bone at the proximal end. A general recommendation made by some is to try and reduce the bone and plate together with the help of pelvic reduction clamps. We have identified this as a source of error in our practice which has a potential to introduce malalignment in tibia, most often discovered post operatively. Here we will describe with relevant cases, the pitfall in the recommended technique and steps to be taken to avoid this error.
机译:微创技术以及新的预轮廓锁定板(LCP PLT,LISS PLT)的广泛使用引起了胫骨近端骨折中许多与对准有关的问题。在胫骨近端骨折中使用预先成型的锁定板所引起的与种植体相关的重大问题是板的上端对皮下组织的硬件刺激。这是由于植入物的较大轮廓引起的,该轮廓位于近端的骨头上。一些人的一般建议是尝试在骨盆复位钳的帮助下减少骨骼和板块。我们认为这是我们实践中的错误根源,有可能在胫骨中引入错位,最常见于术后。在这里,我们将以相关案例描述推荐技术中的陷阱以及为避免此错误而应采取的步骤。