首页> 外文期刊>Applied Physics Letters >Carbon nanotube based microfocus field emission x-ray source for microcomputed tomography
【24h】

Carbon nanotube based microfocus field emission x-ray source for microcomputed tomography

机译:基于碳纳米管的微聚焦层析X射线源

获取原文
获取原文并翻译 | 示例
       

摘要

Microcomputed tomography is now widely used for in vivo small animal imaging for cancer studies. Achieving high imaging quality of live objects requires the x-ray source to have both high spatial and temporal resolutions. Preliminary studies have shown that carbon nanotube (CNT) based field emission x-ray source has significant intrinsic advantages over the conventional thermionic x-ray tube including better temporal resolution and programmability. Here we report the design and characterization of a CNT based field emission x-ray source that also affords a high spatial resolution. The device uses modified asymmetric Einzel lenses for electron focusing and an elliptical shaped CNT cathode patterned by photolithography. Stable and small isotropic x-ray focal spot sizes were obtained.
机译:现在,微计算机断层扫描已广泛用于体内小动物成像研究,以进行癌症研究。为了获得高品质的活物成像,需要X射线源具有高的空间和时间分辨率。初步研究表明,基于碳纳米管(CNT)的场发射X射线源具有优于常规热电子X射线管的显着内在优势,包括更好的时间分辨率和可编程性。在这里,我们报告了基于CNT的场发射X射线源的设计和表征,该X射线源还提供了高空间分辨率。该设备使用改进的非对称Einzel透镜进行电子聚焦,并使用光刻技术将椭圆形CNT阴极图案化。获得了稳定且较小的各向同性X射线焦点尺寸。