首页> 外文期刊>Applied Optics >Fourier-transform cavity-enhanced absorption spectroscopy using an incoherent broadband light source
【24h】

Fourier-transform cavity-enhanced absorption spectroscopy using an incoherent broadband light source

机译:使用非相干宽带光源的傅立叶变换腔增强吸收光谱

获取原文
获取原文并翻译 | 示例
           

摘要

A cavity-enhanced absorption setup employing an incoherent broadband light source was used in combination with a Fourier-transform spectrometer to measure the spin-forbidden B-band of gaseous oxygen at ~688 nm and several weak absorption transitions of water vapor in the same spectral region at room temperature in ambient air. The experiments demonstrate that the sensitivity of a Fourier-transform spectrometer can be significantly improved by increasing the effective path length, while retaining a rather small sample volume. In comparison with a single-pass absorption measurement, we report a path-length enhancement factor of 200 and an improvement of the signal-to-noise ratio of ≈ 6 in the present cavity-enhanced absorption experiment. The practical advantages and limitations of this novel approach are outlined and potential applications are briefly discussed.
机译:使用不相干宽带光源的腔增强吸收装置与傅立叶变换光谱仪结合使用,测量了在〜688 nm处自旋禁止的气态氧B带以及在同一光谱中水蒸气的几个弱吸收跃迁在室温下在环境空气中的区域。实验表明,通过增加有效路径长度,同时保留相当小的样本量,可以显着提高傅里叶变换光谱仪的灵敏度。与单次通过吸收测量相比,我们报告了在当前腔增强吸收实验中,光程增强因子为200,信噪比提高了≈6。概述了这种新颖方法的实际优点和局限性,并简要讨论了潜在的应用。