首页> 外文期刊>Analytical and Bioanalytical Chemistry >Amperometric biosensor for the determination of creatine
【24h】

Amperometric biosensor for the determination of creatine

机译:安培生物传感器测定肌酸

获取原文
获取原文并翻译 | 示例
           

摘要

An amperometric biosensor for the determination of creatine was developed. The carbon rod electrode surface was coated with sarcosine oxidase (SOX) and creatine amidinohydrolase by cross-linking under glutaraldehyde vapour. The SOX from Arthrobacter sp. 1–1 N was purified and previously used for creation of a creatine biosensor. The natural SOX electron acceptor, oxygen, was replaced by an $ {{left[ {Fe{left( {CN} right)}_{6} } right]}^{{3 - }} } mathord{left/ {vphantom {{{left[ {Fe{left( {CN} right)}_{6} } right]}^{{3 - }} } {{left[ {Fe{left( {CN} right)}_{6} } right]}}}} right. kern-0em} {{left[ {Fe{left( {CN} right)}_{6} } right]}}^{{4 - }} $ redox mediating system, which allowed amperometric detection of an analytical signal at +400-mV potential. The response time of the biosensor was less than 1 min. The biosensor showed a linear dependence of the signal vs. creatine concentration at physiological creatine concentration levels. The optimal pH in 0.1 M tris(hydroxymethyl)aminomethane (Tris)–HCl buffer was found to be at pH 8.0. The half-life of the biosensor was 8 days in 0.1 M Tris–HCl buffer (pH 8.0) at 20 °C.
机译:开发了用于测定肌酸的安培生物传感器。在戊二醛蒸汽下通过交联将碳棒电极表面涂上肌氨酸氧化酶(SOX)和肌酸a水解酶。来自节杆菌属的SOX。纯化了1-1 N,并先前用于创建肌酸生物传感器。天然的SOX电子受体氧被$ {{left [{Fe {left({CN} right)} _ {6}} right]} ^ {{3-}}}} mathord {left / {vphantom {{{left [{Fe {left({CN} right)} _ {6}}右]} ^ {{3-}}}} {{left [{Fe {left({CN} right)} _ {6 } } 是的是的。 kern-0em} {{左[{Fe {左({CN}右}} _ {6}}右}}} {{4-}} $氧化还原介导系统,允许在+处安培检测分析信号400-mV电位。生物传感器的响应时间少于1分钟。生物传感器在生理性肌酸浓度水平下显示信号对肌酸浓度的线性依赖性。发现在0.1 M三(羟甲基)氨基甲烷(Tris)-HCl缓冲液中的最佳pH值为8.0。生物传感器在20°C下于0.1 M Tris-HCl缓冲液(pH 8.0)中的半衰期为8天。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号