In vitro performance of a new non-invasive extracorporeal saturation monitor

机译:新型无创体外饱和度监测仪的体外性能

获取原文
获取原文并翻译 | 示例
           

摘要

A simulated cardiopulmonary bypass circuit was constructed in the laboratory to compare the accuracy and precision of a new non-invasive extracorporeal oxygen saturation monitor – the M3 monitor (Spectrum Medical LLP, Gloucester, UK) against the Siemens RAPIDLab? blood gas analyser (Siemens AG, Munich, Germany). Comparisons were made across a range of oxygen saturations and at different temperatures and different haemoglobin concentrations. Results showed that under all conditions, when recording oxygen saturation, the M3 monitor achieved a mean (SD) bias of ?0.9 (1.7)% with limits of clinical agreement of ?4.2 to 2.5. However, the difference between the two monitors was larger at lower saturation levels (p?
机译:在实验室中构建了一个模拟的心肺旁路电路,以比较新型无创体外氧饱和度监测器– M3监测器(Spectrum Medical LLP,英国格洛斯特)与西门子RAPIDLab ?血气分析仪(西门子,德国慕尼黑)。在不同的氧饱和度范围,不同的温度和不同的血红蛋白浓度下进行了比较。结果表明,在所有条件下,当记录氧饱和度时,M3监护仪均达到±0.9(1.7)%的平均(SD)偏倚,且临床一致性限制为±4.2至2.5。然而,在较低的饱和度水平(p≤0.001),较低的血红蛋白水平(p≤0.002)和较低的温度(p = 0.013)下,两个监测器之间的差异更大。用M3监测仪记录的平均(SD)血红蛋白浓度为15.4(0.6),而用血气分析仪测量的平均血红蛋白浓度为15.4(0.5)?g.dl ?1 。血气分析仪的黄金标准,具有连续读取的优点。

著录项

  • 来源
    《Anaesthesia》 |2011年第10期|889-894|共6页
  • 作者单位

    Consultant Anaesthetist Alexandra Hospital Redditch UK;

    Consultant Anaesthetist Royal Orthopaedic Hospital NHS Foundation Trust Birmingham UK;

    Senior Lecturer in Anaesthesia and Intensive Care University Hospitals Birmingham NHS Foundation Trust The New Queen Elizabeth Hospital Birmingham UK;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号