首页> 外文期刊>Anaesthesia >Ability of non-invasive intermittent blood pressure monitoring and a continuous non-invasive arterial pressure monitor (CNAP?) to provide new readings in each 1-min interval during elective caesarean section under spinal anaesthesia
【24h】

Ability of non-invasive intermittent blood pressure monitoring and a continuous non-invasive arterial pressure monitor (CNAP?) to provide new readings in each 1-min interval during elective caesarean section under spinal anaesthesia

机译:脊髓麻醉下选择性剖宫产术中每1分钟间隔内无创间歇性血压监测和连续无创动脉压监测器(CNAP ?)的功能

获取原文
获取原文并翻译 | 示例
       

摘要

We compared the ability of automated non-invasive intermittent oscillometric blood pressure monitoring with a new device, CNAPTM (continuous non-invasive arterial pressure) to provide a new blood pressure reading in each 1-min interval between spinal anaesthesia and delivery during caesarean section. We also compared the accuracy of continuous non-invasive arterial pressure readings with non-invasive blood pressure measurements before spinal anaesthesia. Fifty-nine women participated. The non-invasive and continuous non-invasive monitors displayed new blood pressure readings in a mean of 82% (11%) and 83% (13%) (p = 0.97) of the one-minute intervals between spinal anaesthesia and delivery, respectively. Continuous non-invasive arterial pressure was more likely to fail on two or more consecutive minutes (p?=?0.001). From the pre-spinal readings, the mean bias, defined as non-invasive – continuous non-invasive arterial pressure, and limits of agreement (±2SD mean bias) for systolic, diastolic and mean blood pressure respectively were +1.3 (±26.0), ?2.9 (±21.8) and +2.6 (±20.4) mmHg. The new monitor has disadvantages compared with conventional non-invasive intermittent blood pressure monitoring.
机译:我们比较了自动无创间歇性示波血压监测与新设备CNAP TM (连续无创动脉压)在每1分钟间隔内提供新血压读数的能力剖宫产时的脊柱麻醉和分娩。我们还比较了在麻醉前连续无创动脉压读数与无创血压测量值的准确性。 59名妇女参加了会议。非侵入性和连续性非侵入性监护仪分别在脊髓麻醉和分娩之间的一分钟间隔内分别显示平均82%(11%)和83%(13%)(p = 0.97)的新血压读数。连续的非侵入性动脉压更有可能在连续两分钟或更长时间内失效(p = 0.001)。从脊柱前的读数中,定义为非侵入性–连续非侵入性动脉压的平均偏倚和收缩压,舒张压和平均血压的一致极限(±2SD平均偏倚)分别为+1.3(±26.0) ,≤2.9(±21.8)和+2.6(±20.4)mmHg。与常规的非侵入性间歇性血压监测相比,该新型监测器具有缺点。

著录项

  • 来源
    《Anaesthesia》 |2012年第3期|274-279|共6页
  • 作者单位

    Registrar;

    Consultant Department of Anaesthesia The Royal Women’s Hospital Melbourne Australia;

    Clinical Associate Professor Department of Pharmacology University of Melbourne Australia;