Evaluation of X-ray doses and their corresponding biological effects on experimental animals in cone-beam micro-CT scans (R-mCT2)

机译:在锥形束微CT扫描(R-mCT2)中评估X射线剂量及其对实验动物的相应生物学效应

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Studies show that the radiation dose received during a micro-CT examination may have adverse effects on living subjects. However, the correlations between the biological effects and the radiation doses have never been thoroughly evaluated in the majority of cases. In this study, we evaluated the biological radiation effects of measured radiation doses in ICR mice using cone-beam micro-CT scans. Long-term in vivo whole-body micro-CT scans of ICR mice were performed for a duration of 4 weeks. Although a scanning frequency of three scans per week is higher than that necessary for conventional studies, this study represents particular cases where the subjects may undergo an extreme number of examinations. The average X-ray dose of a CT scan measures 16.19 mGy at the center of a phantom and 16.24 mGy at an offset position of 7.5 mm from the center of the phantom. The total average dose at the center of the phantom during the 4-week scanning period was 194.3 mGy. No significant radiation effects were observed in the weight gain curves, organ weights, blood analyses, litter sizes, reared offspring sizes, and the histopathologic results. Therefore, it is unlikely that the measured doses for the CT scans caused any radiation damage in the mice.
机译:研究表明,在微型CT检查期间接收到的辐射剂量可能会对活着的受试者产生不利影响。然而,在大多数情况下,尚未完全评估生物学效应和辐射剂量之间的相关性。在这项研究中,我们使用锥束微CT扫描评估了ICR小鼠中测得的辐射剂量对生物辐射的影响。对ICR小鼠进行长期体内全身微CT扫描,持续4周。尽管每周三次扫描的扫描频率高于常规研究所必需的频率,但该研究代表了受试者可能需要进行大量检查的特殊情况。 CT扫描的平均X射线剂量在体模中心测得16.19 mGy,在距体模中心7.5 mm的偏移位置测得16.24 mGy。在4周的扫描期间内,幻影中央的总平均剂量为194.3 mGy。在体重增加曲线,器官重量,血液分析,垫料大小,后代大小和组织病理学结果中均未观察到明显的辐射影响。因此,CT扫描的测量剂量不太可能对小鼠造成任何辐射损害。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号