The feasibility of low-concentration contrast and low tube voltage in computed tomography perfusion imaging: an animal study

机译:低浓度造影剂和低管电压在计算机断层扫描灌注成像中的可行性:一项动物研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Aim>: To investigate the feasibility of low-concentration contrast (270 mg/ml) together with low tube voltage (80 kV) and adaptive iterative dose reduction (AIDR)-3D reconstruction in liver computed tomography (CT) perfusion imaging.Method>: A total of 15 healthy New Zealand rabbits received two CT scans each. The first scan (control) was acquired at 100 kV and 100 mA with iopromide (370 mg/ml), while the second scan (experimental) was acquired at 80 kV and 100 mA with iodixanol (270 mg/ml) 24 h after the first scan. The obtained images were reconstructed with filtered back projection (FBP) and AIDR-3D in the control and experimental groups respectively. The perfusion parameters (hepatic artery perfusion [HAP], portal vein perfusion [PVP], hepatic perfusion index [HPI], and total liver perfusion [TLP]) and image quality (image quality score, average CT value of abdomen aorta, signal-to-noise ratio [SNR], contrast-to-noise ratio [CNR], and figure of merit [FOM]) were compared using a paired t-test or Mann–Whitney U test between the two groups, when appropriate. The effective radiation dose and iodine intake were also recorded and compared.Results>: With the exception of the FOM criteria, the image quality and perfusion parameters were not significantly different between the two groups. The effective radiation dose and iodine intake were 38.79% and 27.03% lower respectively, in the experimental group.Conclusion>: Low-concentration contrast (iodixanol, 270 mg/ml) together with low tube voltage (80 kV) and AIDR-3D reconstruction help to reduce radiation dose and iodine intake without compromising perfusion parameters and image quality in liver CT perfusion imaging.
机译:目的>:研究低浓度造影剂(270 mg / ml)以及低管电压(80 kV)和自适应迭代剂量减少(AIDR)-3D重建在肝脏计算机断层扫描(CT)中的可行性方法>::总共15只健康的新西兰兔每只接受了两次CT扫描。第二次扫描(对照)是在碘化丙啶(370 mg / ml)下于100 kV和100 mA下获得的,而第二次扫描(实验)是在24 h后用碘克沙醇(270 mg / ml)在80 kV和100 mA下获得的。初次扫描。分别在对照组和实验组中用滤波反投影(FBP)和AIDR-3D重建获得的图像。灌注参数(肝动脉灌注[HAP],门静脉灌注[PVP],肝灌注指数[HPI]和总肝灌注[TLP])和图像质量(图像质量评分,腹主动脉平均CT值,信号适当时,使用配对t检验或Mann-Whitney U检验对两组之间的信噪比[SNR],对比信噪比[CNR]和品质因数[FOM]进行比较。结果>::除FOM标准外,两组的图像质量和灌注参数无明显差异。实验组的有效放射剂量和碘摄入量分别降低38.79%和27.03%。结论>:低浓度造影剂(碘克沙醇,270 mg / ml)和低管电压(80 kV) )和AIDR-3D重建有助于减少辐射剂量和碘摄入量,而不会损害肝脏CT灌注成像中的灌注参数和图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号