首页> 外文会议>Conference on physics of medical imaging >Material Decomposition and Virtual Non-Contrast Imaging in Photon Counting Computed Tomography - An Animal Study
【24h】

Material Decomposition and Virtual Non-Contrast Imaging in Photon Counting Computed Tomography - An Animal Study

机译:光子计数计算机断层扫描中的材料分解和虚拟非对比成像-动物研究

获取原文

摘要

The energy resolving capabilities of Photon Counting Detectors (PCD) in Computed Tomography (CT) facilitate energy-sensitive measurements. The provided image-information can be processed with Dual Energy and Multi Energy algorithms. A research PCD-CT firstly allows acquiring images with a close to clinical configuration of both the X-ray tube and the CT-detector. In this study, two algorithms (Material Decomposition and Virtual Non-Contrast-imaging (VNC)) are applied on a data set acquired from an anesthetized rabbit scanned using the PCD-CT system. Two contrast agents (CA) are applied: A gadolinium (Gd) based CA used to enhance contrasts for vascular imaging, and xenon (Xe) and air as a CA used to evaluate local ventilation of the animal's lung. Four different images are generated: a) A VNC image, suppressing any traces of the injected Gd imitating a native scan, b) a VNC image with a Gd-image as an overlay, where contrast enhancements in the vascular system are highlighted using colored labels, c) another VNC image with a Xe-image as an overlay, and d) a 3D rendered image of the animal's lung, filled with Xe, indicating local ventilation characteristics. All images are generated from two images based on energy bin information. It is shown that a modified version of a commercially available dual energy software framework is capable of providing images with diagnostic value obtained from the research PCD-CT system.
机译:计算机断层扫描(CT)中的光子计数检测器(PCD)的能量分辨能力有助于进行能量敏感的测量。所提供的图像信息可以使用双能量和多能量算法进行处理。一项研究型PCD-CT首先允许以接近X射线管和CT探测器的临床配置的方式获取图像。在这项研究中,两种算法(材料分解和虚拟非对比成像(VNC))应用于从使用PCD-CT系统扫描的麻醉兔子获取的数据集。应用了两种造影剂(CA):基于g(Gd)的CA用于增强血管成像的对比度,而氙(Xe)和空气作为CA用于评估动物肺部的局部通气。生成四个不同的图像:a)VNC图像,模拟自然扫描,抑制了注入的Gd的任何痕迹,b)以Gd图像为覆盖的VNC图像,其中使用彩色标签突出显示了血管系统的对比度增强; c)另一个以Xe图像作为覆盖的VNC图像,以及d)用Xe填充的动物肺部3D渲染图像,表示局部通气特征。所有图像都是基于能量仓信息从两个图像生成的。结果表明,市售双能软件框架的修改版能够提供具有从研究PCD-CT系统获得的诊断价值的图像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号