首页> 美国卫生研究院文献>The Journal of Neurology and Psychopathology >Laboratory testing of the Spiegelberg brain pressure monitor: atechnical report
【2h】

Laboratory testing of the Spiegelberg brain pressure monitor: atechnical report

机译:Spiegelberg脑压监测器的实验室测试:技术报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVES—The Spiegelberg brain pressure monitor is a low cost intracranial pressure monitoring system that has been used clinically for some time, mainly in Germany. To provide a rigorous bench comparison of the Spiegelberg monitor with the Camino pressure monitor an evaluation programme has been carried out in the UK Shunt Evaluation Laboratory.
DESIGN—Drift over 72 hours and with temperature, a frequency response, and the accuracy of measurement of both static and pulsatile pressures have been tested simultanously in Camino and Spiegelberg transducers using a computerised rig.
RESULTS—Long term zero drift was less than 0.7 mm Hg in both transducers. The Spiegelberg monitor showed no temperature drift whereas the Camino monitor had a drift of around 0.3 mm Hg/°C. The Spiegelberg monitor underread mean pressures <40 mm Hg by <1 mm Hg, but the error increased to 4.7 mm Hg at 100 mm Hg. The frequency bandwidth of the Spiegelberg monitor was 4 Hz at a low pressure. Underreading of the amplitude increased with the mean pressure, with a delay of about 0.1 s in the detection of the peaks of pulse waveform.
CONCLUSION—The Spiegelberg transducer had excellent accuracy for static intracranial pressure measurement, butcomplex waveform analysis may be biased by its limited dynamic response.

机译:目的-Spiegelberg脑压监测器是一种低成本的颅内压监测系统,已在临床上使用了一段时间,主要在德国。为了对Spiegelberg监护仪和Camino压力监护仪进行严格的台架比较,英国分流评估实验室已实施了一项评估计划。
设计—漂移超过72小时,并具有温度,频率响应和精度在Camino和Spiegelberg换能器中,同时使用计算机装置同时测试了静压和脉动压的测量结果。
结果-两种换能器的长期零漂移均小于0.7 mm Hg。 Spiegelberg监测器没有温度漂移,而Camino监测器的漂移约为0.3 mm Hg /°C。 Spiegelberg监测仪的平均压力低于40 mm Hg,小于1 mm Hg,但在100 mm Hg时,误差增加到4.7 mm Hg。 Spiegelberg监护仪的频率带宽在低压下为4 Hz。振幅的欠读随着平均压力的增加而增加,在检测脉冲波形的峰值时会延迟约0.1 s。
结论— Spiegelberg传感器在静态颅内压测量中具有出色的精度,但是复杂的波形分析可能会因其有限的动态响应而产生偏差。