首页> 美国卫生研究院文献>The Journal of Automatic Chemistry >Automated cold vapour flow-injection analysis of mercury at high concentrations
【2h】

Automated cold vapour flow-injection analysis of mercury at high concentrations

机译:高浓度汞的自动冷蒸气流进样分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Continuous-flow cold vapour- atomic fluorescence spectrometry is shown to be an extremely sensitive technique for the determination of mercury with detection limits typically below 0.01 μg l-1. Linear calibration ranges were found to be at least four orders of magnitude (i.e. up to 0.1 mg l-1). Samples with concentrations exceeding the linear range are susceptible to self-absorption, and may, in severe cases, cause carry-over problems between samples. The flow-injection approach has been utilized to extend the upper limit of the linear calibration range allowing determinations up to 10 mg l-1 of mercury. A range of certified reference materials and zinc battery anodes have been successfully analysed with a minimal number of sample dilutions.
机译:连续流冷蒸气原子荧光光谱法是一种非常灵敏的测定汞的技术,检测限通常低于0.01μgl -1 。发现线性校准范围至少为四个数量级(即最高0.1 mg l -1 )。浓度超出线性范围的样品容易吸收,在严重的情况下可能会导致样品之间残留问题。流动注射法已被用于扩展线性校准范围的上限,从而可以测定高达10 mg l -1 的汞。使用最少数量的样品稀释液就可以成功地分析多种认证的参考材料和锌电池负极。