首页> 美国卫生研究院文献>Gut >Electrodes for 24 hours pH monitoring--a comparative study.
【2h】

Electrodes for 24 hours pH monitoring--a comparative study.

机译:用于24小时pH监测的电极-一项对比研究。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

Three pH electrodes in clinical use were examined--(1) antimony electrode with remote reference electrode (Synectics 0011), (2) glass electrode with remote reference electrode (Microelectrodes Inc. MI 506) and (3) combined glass electrode with integral reference electrode (Radiometer GK2801C). In vitro studies showed that both glass electrodes were similar and superior to the antimony electrode with respect to response time, drift, and sensitivity. The effect of the siting of the reference electrode on the recorded pH was examined in five human volunteers. The pH reading using a remote skin reference electrode was higher by a mean of 0.3 pH units (range 0.0-0.6) in the stomach, lower by 0.65 pH units (0.5-0.8) in the duodenum and lower by 0.3 pH units (0.0-0.6) in the oesophagus than that simultaneously obtained with an intraluminal reference electrode. Buccal reference electrodes gave similar readings to skin. Combined reference and glass pH electrodes are recommended for 24-hour ambulatory pH monitoring.
机译:检查了三个临床使用的pH电极-(1)带有远程参比电极的锑电极(Synectics 0011),(2)带有远程参比电极的玻璃电极(Microelectrodes Inc. MI 506)和(3)带有整体参比的组合玻璃电极电极(辐射计GK2801C)。体外研究表明,两个玻璃电极在响应时间,漂移和灵敏度方面均相似且优于锑电极。在五名人类志愿者中检查了参比电极的位置对记录的pH值的影响。使用远端皮肤参比电极的pH值在胃中平均升高0.3个pH单位(范围0.0-0.6),在十二指肠中降低0.65个pH单位(0.5-0.8),而在胃中降低0.3个pH单位(0.0-0.0)。在食管中比在腔内参比电极上同时获得的高0.6)。颊部参比电极对皮肤的读数相似。建议将参比电极和玻璃pH电极组合使用,以进行24小时动态pH监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号