摘要

>Purpose>: To investigate whether commonly used test tube warming devices maintain a constant temperature in follicular fluid aspirates.>Methods>: By using a digital thermocouple, temperature was measured and comparisons were made between an analog dry block heater, a digital dry block heater, and a thermostatic test tube heater.>Results>: For small fluid volumes, temperature in the block heaters increased above 37°C after being in the block for over 2 min. The thermostatic heater maintained a constant temperature, but this was below the factory setting of 36.9°C. Temperature maintenance was influenced by fluid volume in each tube.>Conclusions>: One of the key factors in the handling of gametes and embryos is the maintenance of constant temperature. Test tube warming devices require verification of their ability to maintain fluid at the desired temperature. Temperature may vary with fluid volume and the type of test tube warming device used.
机译:>目的 >::调查常用的试管加热装置在卵泡液抽吸物中是否保持恒温。>方法 >:通过使用数字热电偶,测量温度并在模拟干式加热器,数字式干式加热器和恒温试管加热器之间进行比较。>结果 >:当流体体积较小时,在块中放置2分钟以上后,块加热器中的温度升高到37°C以上。恒温加热器保持恒定温度,但低于工厂设定的36.9°C。温度维持受每个试管中液体量的影响。>结论 >:维持配子和胚胎的关键因素之一是维持恒温。试管加热装置需要验证其将流体保持在所需温度的能力。温度可能随流体量和所用试管加热装置的类型而变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号