MR Ventriculocisternography By Using 3D Balanced Steady-State Free Precession Imaging: Technical Note

机译:Undiculocisternography先生使用3D平衡稳态免费预防成像:技术说明

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Summary: This study investigated the effects of flip angle setting in 3D balanced steady-state free precession (SSFP) imaging on CSF-parenchyma contrast and section aliasing artifacts. Theoretical derivations indicated that the extent of section aliasing artifacts decreased as the flip angle was lowered, at the expense of a sacrifice in CSF-parenchyma contrast. Experimental data agreed closely with theoretical predictions. A flip angle of about 40° is therefore recommended for 3D balanced SSFP MR ventriculocisternography.
机译:摘要:本研究调查了翻转角度设定在CSF-实质对比度和叠片术语上的3D平衡稳态自由进样(SSFP)成像中的影响。理论衍生表明,随着翻转角度降低的锯齿术曲面的程度,以CSF-实质对比度为牺牲。实验数据与理论预测密切一致。因此建议为3D平衡SSFP MR Ventriculocistralography建议使用约40°的翻转角。