首页> 美国卫生研究院文献>Journal of Neurological Surgery Reports >Dynamic Occlusion of Distal Ventriculoperitoneal Shunt Catheter after Infusion Port Placement: A New Shunt Malfunction
【2h】

Dynamic Occlusion of Distal Ventriculoperitoneal Shunt Catheter after Infusion Port Placement: A New Shunt Malfunction

机译:输注端口放置后远端脑室分流管的动态闭塞:新的分流器故障

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Shunt failure requiring reintervention remains a common complication of hydrocephalus treatment. Here, we report a novel cause of mechanical shunt obstruction in an adult patient: position-dependent intermittent occlusion via an infusion port catheter. A 51-year-old woman with a grade II oligodendroglioma presented in a delayed fashion following surgery with a pseudomeningocele. She underwent ventriculoperitoneal shunt placement due to communicating hydrocephalus, resolving her pseudomeningocele. Shortly thereafter, she underwent placement of a subclavian infusion port at an outside institution. Her pseudomeningocele returned. Imaging demonstrated close proximity of her port catheter to the shunt catheter overlying the clavicle. Her shunt was tapped demonstrating a patent ventricular catheter with normal pressure. She underwent shunt exploration after her pseudomeningocele did not respond to valve adjustment. Intraoperative manometry demonstrated head position-dependent distal catheter obstruction. Repeat manometry following distal catheter revision demonstrated normal runoff independent of position. Her pseudomeningocele was resolved on follow-up. To our knowledge, this is the only reported case of intermittent, position-dependent distal catheter obstruction. Shunted patients with concern for malfunction following subclavian infusion port placement should be evaluated for possible dynamic obstruction of their distal catheter when the two catheters are in close proximity along the clavicle.
机译:需要重复的分流失败仍然是脑积水治疗的常见并发症。在这里,我们在成年患者中报告了一种新颖的机械分流阻塞的原因:通过输注端口导管依赖于位置依赖性间歇闭塞。一名51岁的女性,患有延迟的时尚患者患有延迟的时尚,伴有假表征诱饵。由于沟通脑积水,她接受了脑室分流分流安置,解决了她的假细胞细胞。此后不久,她在外部机构安置了锁骨头输注港。她的假影窝回来了。成像表明她的端口导管靠近覆盖锁骨的分流导管。仔细挖掘了她的分流,证明具有常压的专利性心室导管。在她的Pseudomeningocele没有响应阀门调整后,她经历了分流探索。术中测压表现出头部位置依赖性远端导管障碍物。远端导管修订后的重复测压表明普通径流无关。她的假影窝在后续行动中得到了解决。据我们所知,这是唯一报告的间歇性,位置依赖性远端导管阻塞。应评估亚克拉夫输注端口放置后的故障担忧的患者,当两条导管沿锁骨近距离接近时,应评估其远端导管的动态阻塞。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号