Ultra-trace determination of oxyhalides in ozonated aquacultural marine waters by direct injection ion chromatography coupled with triple-quadrupole mass spectrometry

机译:通过直接喷射离子色谱法与三腹极质谱耦合含臭味的水上养殖水域氧化物的超痕量测定

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A direct, robust, accurate and highly sensitive method for oxyhalide species in natural waters, including seawater, using suppressed ion chromatography coupled with mass spectrometry (IC-MS) is described. The method utilised a high capacity, high efficiency anion-exchange column (Dionex IonPac AS11-HC, 4 mm, 2 × 250 mm), with the separation achieved using an electrolytically generated potassium hydroxide gradient, delivered at 0.380 mL min−1. Applying the method, detection limits for iodate, bromate, and chlorate in seawater after direct sample injection (20 μL injection volume, samples diluted 10-fold), were 11, 30 and 13 ng L−1 (ppt), respectively. Standard addition calibrations to ozonated seawater samples were linear, in all cases R2 > 0.999 (n = 10), with intra-day repeatability of 3.7, 11.2 and 1.8 % RSD (n = 10) for a low-level standard mixture (0.30 μg L−1 of iodate, 0.15 μg L−1 of bromate, and 1.50 μg L−1 of chlorate). The method was applied to the analysis of seawater samples taken pre- and post-disinfection points within a recirculating aquacultural system. Iodate, bromate and chlorate were detected as the main oxyanionic disinfection by-products, demonstrating the practical utility of the new method as a valuable tool for monitoring changes to seawater composition following disinfection treatments.
机译:描述了使用抑制离子色谱(包括海水,与质谱(IC-MS)偶联的天然水域中的氧化氧化物物种的直接,鲁棒,准确和高敏感的方法。该方法利用高容量,高效阴离子交换柱(Dionex Ionpac As11-Hc,4mm,2×250mm),使用电解产生的氢氧化钾梯度实现的分离,在0.380mL min-1中递送。在直接样品注射(20μL注射体积,10倍)的20μL注射体积,样品)之后,将碘酸盐,溴酸盐和氯酸盐施加方法,溴酸盐和氯酸盐分别为11,30和13ng L-1(PPT)。向臭氧海水样品的标准加法校准是线性的,在所有情况下,r2> 0.999(n = 10),具有3.7,11.2和1.8%RSD(N = 10)的日期重复性(0.30μg L-1碘酸盐,0.15μgL-1溴酸盐,1.50μgL-1的氯酸盐)。该方法应用于海水样本的分析,在循环水产系统内采取的海水样本和消毒后斑点。碘化物,溴酸盐和氯酸盐被检测为主要的氧气消毒副产物,展示了新方法的实用效用作为在消毒处理后监测海水组合物的有价值的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号