An evaluation of systematic errors on marker-based registration of computed tomography and magnetic resonance images of the liver

机译:对肝脏计算断层扫描和磁共振图像的基于标记的基于标记的系统误差的评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We demonstrated a general method to evaluate systematic errors related to Magnetic Resonance (MR) imaging sequences in marker-based co-registration of MR and Computed Tomography (CT) images, and investigated the effect of MR image quality in the co-registration process using clinical MR and CT protocols for stereotactic ablative body radiotherapy (SABR) planning of the liver. Small systematic errors (under 1.6 mm) were detected, unlikely to be a clinical risk to liver SABR. The least favourable marker configuration was found to be a co-planar arrangement parallel to the transaxial image plane.
机译:我们展示了一种评估与MR和计算机断层扫描(CT)图像的标记的共同登记中的磁共振(MR)成像序列相关的系统误差,并研究了使用的共同登记过程中MR图像质量的影响肝脏立体定向烧蚀体放射治疗(SABR)规划的临床MR和CT协议。检测到肝脏SABR的临床风险,检测到小型系统误差(下降1.6毫米)。发现最不利的标记配置是平行于颠膜图像平面的共平面布置。