Component fracture is a rare cause for revision in total hip arthroplasty. For a fractured well-fixed long femoral stem, the options are limited. We sought to develop a technique to address this problem with lower morbidity. A newly developed cemented tube was constructed and cemented onto a fractured Revitan revision hip femoral stem to retain the distal well-fixed component. At the 2-year follow-up, the Harris Hip Score, pain level, and radiographic images were analyzed. At the 2-year follow-up, no radiological signs of loosening or failure could be observed. The patient’s preoperative Harris Hip Score improved from 42.8 to 97 points. The pain level improved from 7/10 to 0/10. Our case report depicts excellent clinical and radiographic outcomes at 2-year follow-up by a newly developed cemented tube technique. This technique is a potential new option for revision of fractured well-fixed diaphyseal stems without major bone loss. Our successful results suggest this technique is worthy of consideration and further study.
机译:组分骨折是在总髋关节置换术中修复的罕见原因。对于裂缝固定的长股骨杆,选项有限。我们试图开发一种解决这种问题,以解决较低的发病率。将新开发的粘合管构造并粘合在骨折的Revitan修正髋关节杆上,以保持远端固定部件。在2年的随访中,分析了哈里斯臀部得分,疼痛水平和放射线图像。在为期两年的随访中,无法观察到宽松或失败的放射性迹象。患者的术前哈里斯臀部分数从42.8%提高到97分。疼痛水平从7/10到0/10提高。我们的案例报告描绘了通过新开发的粘液管技术在2年的后续行动中描述了出色的临床和射线照相结果。这种技术是用于修改破裂良好固定的透析性茎的潜在的新选择,没有重大骨质损失。我们的成功结果表明这种技术值得考虑和进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号