首页> 美国卫生研究院文献>Journal of Vascular Surgery Cases and Innovative Techniques >The two-cut monorail technique for the over-the-wire removal of the Impella CP device
【2h】

The two-cut monorail technique for the over-the-wire removal of the Impella CP device

机译:双裁剪单轨技术用于偶尔剥离偶极疱疹CP装置

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Impella is a percutaneously placed intra-arterial flow pump positioned across the aortic valve for circulatory support. A limitation of the Impella is that it lacks a central wire channel, to maintain intra-arterial wire access when removing the device. Open surgical arterial cutdown is needed for the removal of the Impella CP placed emergently, without the use of preclose sutures. This case review describes an alternative removal method for the aforementioned occasions.
机译:偶像是经皮下放置在主动脉瓣膜的动脉内流动泵,用于循环支撑。偶像的限制是它缺乏中央丝通道,以在去除装置时保持动脉内的线路。不需要使用预制缝合线,所需的移除普拉氏CP放置的普通手术动脉降调。本例审查描述了上述场合的替代删除方法。