The Effects of Cyclic Loading and Motion on the Implant–Cement Interface and Cement Mantle of PEEK and Cobalt–Chromium Femoral Total Knee Arthroplasty Implants: A Preliminary Study

机译:循环载荷和运动对PEEK和钴 - 铬股骨膝关节置换植入物植入水泥界面和水泥地幔的影响:初步研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study investigated the fixation of a cemented PEEK femoral TKA component. PEEK and CoCr implants were subjected to a walking gait cycle for 10 million cycles (MC), 100,000 cycles or 0 cycles (unloaded control). A method was developed to assess the fixation at the cement–implant interface, which exposed the implants to a fluorescent penetrant dye solution. The lateral condyles of the implants were then sectioned and viewed under fluorescence to investigate bonding at the cement–implant interface and cracking of the cement mantle. When tested for 100,000 cycles, debonding of the cement–implant interface occurred in both PEEK (61%) and CoCr (13%) implants. When the duration of testing was extended (10 MC), the percentage debonding was further increased for both materials to 88% and 61% for PEEK and CoCr, respectively. The unloaded PEEK specimens were 79% debonded, which suggests that, when PEEK femoral components are cemented, complete bonding may never occur. Analysis of cracks in the cement mantle showed an absence of full-thickness cracks in the unloaded control group. For the 100,000-cycle samples, on average, 1.3 and 0.7 cracks were observed for PEEK and CoCr specimens, respectively. After 10 MC, these increased to 24 for PEEK and 19 for CoCr. This was a preliminary study with a limited number of samples investigated, but shows that, after 10 MC under a walking gait, substantial debonding was visible for both PEEK and CoCr implants at the cement–implant interface and no significant difference in the number of cement cracks was found between the two materials.
机译:本研究研究了胶合PEEK股骨头TKA组分的固定。 PEEK和COCR植入物受到步行步态循环1000万个循环(MC),100,000个循环或0个循环(卸载控制)。开发了一种方法以评估水泥植入界面处的固定,使植入物暴露于荧光渗透染料溶液。然后在荧光下切割植入物的侧髁,以在水泥 - 植入界面处研究键合和水泥地幔的开裂。当测试100,000次循环时,PEEK(61%)和COCR(13%)植入物中发生了水泥植入界面的剥离。当测试的持续时间延长(10 MC)时,剥离的百分比较低,分别为PEEK和COOC的88%和61%。卸载的PEEK标本达到79%剥夺,这表明,当PEEK股骨成分粘合时,可能永远不会发生完全粘合。水泥地幔中裂缝分析显示卸载对照组中的全厚裂缝。对于100,000次循环样品,平均分别观察到PEEK和COCR标本的1.3和0.7裂缝。 10 MC之后,这些增加到了24个,为Peek和19岁。这是一种初步研究,其中有有限数量的样品,但表明,在步态步态下10mc之后,PEEK和COCR植入水泥植入界面的大量剥离可见,并且水泥数量没有显着差异两种材料之间发现了裂缝。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号