We observed that patients treated with continuous vecuronium or esmolol infusions showed elevated plasma sodium measurements when measured by the routine chemistry analyzer as part of the basic metabolic panel (Vitros 5600; Ortho Clinical Diagnostics, Raritan, NJ), but not by blood gas analyzers (RAPIDLab 1265; Siemens, Tarrytown, NY). Both instruments use direct ion-selective electrode technology, albeit with different sodium ionophores (basic metabolic panel: methyl monensin, blood gas: glass). We questioned if the basic metabolic panel hypernatremia represents artefactual pseudohypernatremia.
机译:我们观察到接受连续维库溴铵或艾司洛尔输注治疗的患者通过常规化学分析仪作为基本代谢指标组的一部分(Vitros 5600; Ortho Clinical Diagnostics,Raritan,NJ)进行测量但血浆血钠分析仪( RAPIDLab 1265;西门子,纽约州塔里敦)。两种仪器都使用直接离子选择电极技术,尽管具有不同的钠离子载体(基本代谢组:甲基莫能菌素,血气:玻璃)。我们质疑基本代谢组高钠血症是否代表人工假性高钠血症。