Discrepancies between cardiovascular magnetic resonance and Doppler echocardiography in the measurement of transvalvular gradient in aortic stenosis: the effect of flow vorticity

机译:心血管磁共振和多普勒超声心动图在主动脉瓣狭窄经瓣膜梯度测量中的差异:流动涡度的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundValve effective orifice area EOA and transvalvular mean pressure gradient (MPG) are the most frequently used parameters to assess aortic stenosis (AS) severity. However, MPG measured by cardiovascular magnetic resonance (CMR) may differ from the one measured by transthoracic Doppler-echocardiography (TTE). The objectives of this study were: 1) to identify the factors responsible for the MPG measurement discrepancies by CMR versus TTE in AS patients; 2) to investigate the effect of flow vorticity on AS severity assessment by CMR; and 3) to evaluate two models reconciling MPG discrepancies between CMR/TTE measurements.
机译:背景瓣膜有效口面积EOA和瓣膜平均压力梯度(MPG)是评估主动脉瓣狭窄(AS)严重程度的最常用参数。但是,通过心血管磁共振(CMR)测量的MPG可能与经胸多普勒超声心动图(TTE)测量的MPG不同。这项研究的目的是:1)找出导致AS患者CMR与TTE的MPG测量差异的因素; 2)研究流动涡度对CMR对AS严重性评估的影响;和3)评估两个模型,以协调CMR / TTE测量之间的MPG差异。