首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Analysis of the Comparative Workflow and Performance Characteristics of the VITEK 2 and Phoenix Systems
【2h】

Analysis of the Comparative Workflow and Performance Characteristics of the VITEK 2 and Phoenix Systems

机译:VITEK 2和Phoenix系统的比较工作流程和性能特征分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

The VITEK 2 (bioMérieux, Marcy L′Ètoile, France) and the Phoenix systems (BD Diagnostic Systems, Sparks, Md.) are automated instruments for rapid organism identification and susceptibility testing. We evaluated the workflow, the time to result, and the performance of identification and susceptibility testing of both instruments. A total of 307 fresh clinical isolates were tested: 141 Enterobacteriaceae, 22 nonfermenters, 93 Staphylococcus spp., and 51 Enterococcus spp. Manipulation time was measured in batches, each with seven isolates, for a total of 39 batches. The mean (± standard deviation [SD]) manipulation time per batch was 20.9 ± 1.8 min for Phoenix and 10.6 ± 1.0 min for VITEK 2 (P < 0.001). Mean (±SD) time to result for all bacterial groups was 727 ± 162 min for Phoenix and 506 ± 120 min for VITEK 2 (P < 0.001). Concerning identification, Phoenix and VITEK 2 yielded the same results for nonfermenters (100%), staphylococci (97%), and enterococci (100%). For 140 Enterobacteriaceae strains evaluated, 135 (96%) were correctly identified by Phoenix and 137 (98%) by VITEK 2 (P = 0.72). The overall category agreement for all isolates was 97.0% for both instruments. The minor error rate, major error rate, and very major error rate for all bacterial isolates tested were 3.0, 0.3, and 0.6 and 2.8, 0.2, and 1.7 for Phoenix and VITEK 2, respectively (P values of 0.76, 0.75, and 0.09). The VITEK 2 system required less manual manipulation time and less time than the Phoenix system to yield results.
机译:VITEK 2(法国Marcy L'Ètoile的bioMérieux)和Phoenix系统(马里兰州Sparks的BD Diagnostic Systems)是用于快速生物识别和药敏测试的自动化仪器。我们评估了工作流程,得出结果的时间以及两种仪器的鉴定和药敏测试的性能。总共测试了307个新鲜的临床分离株:141个肠杆菌科,22个非发酵菌,93个葡萄球菌和51个肠球菌。分批测量操作时间,每批有七个分离物,共计39个批次。 Phoenix的每批次平均操作时间(±标准偏差[SD])为20.9±1.8分钟,VITEK 2的操作时间为10.6±1.0分钟(P <0.001)。所有细菌组的平均(±SD)时间对于Phoenix来说是727±162分钟,对于VITEK 2是506±120分钟(P <0.001)。关于鉴定,Phoenix和VITEK 2对非发酵菌(100%),葡萄球菌(97%)和肠球菌(100%)的结果相同。对于140个评估的肠杆菌科菌株,Phoenix正确鉴定了135个菌株(96%),VITEK 2正确鉴定了137个菌株(98%)(P = 0.72)。两种仪器的所有分离物的总体分类一致性为97.0%。 Phoenix和VITEK 2的所有细菌分离物的次要错误率,主要错误率和非常主要的错误率分别为3.0、0.3和0.6、2.8、0.2和1.7(P值分别为0.76、0.75和0.09) )。与Phoenix系统相比,VITEK 2系统需要更少的手动操作时间和更少的时间来产生结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号