摘要

Four hundred sixty-four fecal specimens were tested for rotavirus by three immunoassays, VIDAS Rotavirus (bioMérieux Vitek, Hazelwood, Mo.), Rotaclone (Cambridge Biotech Corporation, Worcester, Mass.), and Pathfinder Rotavirus (Sansfi Diagnostics Pasteur, Chaska, Minn.). Twenty-seven discordant specimens were evaluated by electron microscopy. The sensitivities, specificities, positive predictive values, and negative predictive values were 98, 99.3, 98.7, and 99%, respectively, for VIDAS Rotavirus; 100, 99, 98.1, and 100%, respectively, for Rotaclone; and 100, 92.4, 87.3, and 100%, respectively for Pathfinder Rotavirus.
机译:通过三种免疫测定法对644份粪便标本进行轮状病毒检测:VIDAS轮状病毒(密苏里州黑泽伍德市的BioMérieuxVitek),Rotaclone(马萨诸塞州伍斯特市的剑桥生物技术公司)和探路者轮状病毒(Sansfi Diagnostics Pasteur,Chaska,明尼苏达州) )。通过电子显微镜评估了二十七个不一致的标本。 VIDAS轮状病毒的敏感性,特异性,阳性预测值和阴性预测值分别为98%,99.3、98.7和99%。 Rotaclone分别为100、99、98.1和100%;以及Pathfinder Rotavirus的100、92.4、87.3和100%。