Biomechanical properties of soft tissues are important for a wide range of medical applications, such as surgical simulation and planning and detection of lesions by elasticity imaging modalities. Currently, the data in the literature is limited and conflicting. Furthermore, to assess the biomechanical properties of living tissue in vivo, reliable imaging-based estimators must be developed and verified. For these reasons we developed and compared two independent quantitative methods – crawling wave estimator (CRE) and mechanical measurement (MM) for soft tissue characterization. The CRE method images shear wave interference patterns from which the shear wave velocity can be determined and hence the Young’s modulus can be obtained. The MM method provides the complex Young’s modulus of the soft tissue from which both elastic and viscous behavior can be extracted. This article presents the systematic comparison between these two techniques on the measurement of gelatin phantom, veal liver, thermal-treated veal liver, and human prostate. It was observed that the Young’s moduli of liver and prostate tissues slightly increase with frequency. The experimental results of the two methods are highly congruent, suggesting CRE and MM methods can be reliably used to investigate viscoelastic properties of other soft tissues, with CRE having the advantages of operating in nearly real time and in situ.
机译:软组织的生物力学特性对于广泛的医学应用非常重要,例如外科手术模拟以及通过弹性成像方式计划和检测病变。当前,文献中的数据有限且相互矛盾。此外,为了评估体内活体组织的生物力学特性,必须开发和验证可靠的基于成像的估计器。由于这些原因,我们开发并比较了两种独立的定量方法-爬行波估计器(CRE)和机械测量(MM)用于软组织表征。 CRE方法对剪切波干涉图进行成像,由此可以确定剪切波速度,从而获得杨氏模量。 MM方法提供了软组织的复杂杨氏模量,从中可以提取弹性和粘性行为。本文介绍了这两种技术在明胶体模,小牛肉肝,热处理小牛肉肝和人前列腺测量中的系统比较。据观察,肝脏和前列腺组织的杨氏模量随频率而略有增加。两种方法的实验结果高度一致,这表明CRE和MM方法可以可靠地用于研究其他软组织的粘弹性质,而CRE的优点是可以实时和就地操作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号