Comparison of the Gono-Pak system with the candle extinction jar for recovery of Neisseria gonorrhoeae.

机译:将Gono-Pak系统与蜡烛消光罐用于恢复淋病奈瑟氏球菌的比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Maximum CO2 levels in the Gono-Pak system were obtained 2 h after activation, with a mean value of 1.5%. Although this is less than the 2 to 3% CO2 level obtained with the candle jar, Gono-Pak produced comparable recovery of Neisseria gonorrhoeae in 357 clinical specimens and stock cultures. Using adjustable CO2 incubators, we found the recovery at 1% CO2 to be comparable to that at 5%, whereas 10% CO2 was inhibitory.
机译:激活后2 h,获得了Gono-Pak系统中的最大CO2水平,平均值为1.5%。尽管这低于用蜡烛罐获得的2%至3%的二氧化碳水平,但Gono-Pak在357个临床标本和原种培养物中产生了与淋病奈瑟菌相当的回收率。使用可调式CO2培养箱,我们发现1%CO2的回收率与5%的回收率相当,而10%CO2的回收率具有抑制性。