Refractive Index Measurement of the Isolated Crystalline Lens Using Optical Coherence Tomography

机译:光学相干断层扫描法测量孤立晶体透镜的折射率

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An optical coherence tomography system has been developed that was designed specifically for imaging the isolated crystalline lens. Cross-sectional OCT images were recorded on 40 lenses from 32 human donors with an age range of 6 – 82 years. A method has been developed to measure the axial thickness and average refractive index of the lens from a single recorded image. The measured average group refractive index at the measurement wavelength of 825 nm was converted to the average phase refractive index at 589 nm using lens dispersion data from the literature. The average refractive index for all lenses measured was 1.408 ± 0.005 which agrees well with recent MRI measurements of the lens index gradient. A linear regression of the data resulted in a statistically significant decrease in the average refractive index with age, but a simple linear model was insufficient to explain the age dependence. The results presented here suggest that the peak refractive index in the nucleus is closer to 1.420, rather than the previously accepted value of 1.406.
机译:已经开发了光学相干断层扫描系统,该系统专门设计用于对孤立的晶状体成像。 OCT横截面图像记录在32位年龄为6至82岁的32位人类捐献者的40片晶状体上。已经开发出一种从单个记录的图像测量透镜的轴向厚度和平均折射率的方法。使用来自文献的透镜色散数据,将在825nm的测量波长处的所测量的平均基团折射率转换为在589nm处的平均相折射率。测量的所有镜片的平均折射率为1.408±0.005,与最近MRI对镜片折射率梯度的测量结果非常吻合。数据的线性回归导致平均折射率随年龄的增长在统计上显着降低,但是简单的线性模型不足以解释年龄的依赖性。此处显示的结果表明,原子核中的峰值折射率接近1.420,而不是先前接受的1.406值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号