Retinal nerve fiber layer (RNFL) thickness, a measure of glaucoma progression, can be measured in images acquired by spectral domain optical coherence tomography (OCT). The accuracy of RNFL thickness estimation, however, is affected by the quality of the OCT images. In this paper, a new parameter, signal deviation (SD), which is based on the standard deviation of the intensities in OCT images, is introduced for objective assessment of OCT image quality. Two other objective assessment parameters, signal to noise ratio (SNR) and signal strength (SS), are also calculated for each OCT image. The results of the objective assessment are compared with subjective assessment. In the subjective assessment, one OCT expert graded the image quality according to a three-level scale (good, fair, and poor). The OCT B-scan images of the retina from six subjects are evaluated by both objective and subjective assessment. From the comparison, we demonstrate that the objective assessment successfully differentiates between the acceptable quality images (good and fair images) and poor quality OCT images as graded by OCT experts. We evaluate the performance of the objective assessment under different quality assessment parameters and demonstrate that SD is the best at distinguishing between fair and good quality images. The accuracy of RNFL thickness estimation is improved significantly after poor quality OCT images are rejected by automated objective assessment using the SD, SNR, and SS.
机译:视网膜神经纤维层(RNFL)的厚度(衡量青光眼进展的程度)可以在通过光谱域光学相干断层扫描(OCT)获取的图像中进行测量。但是,RNFL厚度估计的准确性受OCT图像质量的影响。在本文中,基于OCT图像强度标准偏差的新参数,即信号偏差(SD)被引入到OCT图像质量的客观评估中。还为每个OCT图像计算了其他两个客观评估参数,即信噪比(SNR)和信号强度(SS)。将客观评估的结果与主观评估进行比较。在主观评估中,一名OCT专家根据三个等级(好,中等和差)对图像质量进行了分级。通过客观和主观评估来评估来自六个受试者的视网膜的OCT B扫描图像。通过比较,我们证明了客观评估成功地将可接受的质量图像(好图像和公平图像)和质量较差的OCT图像(由OCT专家分级)区分开来。我们评估了在不同质量评估参数下的客观评估的性能,并证明了SD是区分公平和高质量图像的最佳选择。在使用SD,SNR和SS进行自动客观评估以排除质量较差的OCT图像后,RNFL厚度估计的准确性得到了显着提高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号