首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Laboratory evaluation of five assay methods for vancomycin: bioassay high-pressure liquid chromatography fluorescence polarization immunoassay radioimmunoassay and fluorescence immunoassay.
【2h】

Laboratory evaluation of five assay methods for vancomycin: bioassay high-pressure liquid chromatography fluorescence polarization immunoassay radioimmunoassay and fluorescence immunoassay.

机译:对万古霉素的五种测定方法的实验室评估:生物测定高压液相色谱荧光偏振免疫测定放射免疫测定和荧光免疫测定。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We compared the precision and accuracy of five methods used to measure the concentration of vancomycin in serum: bioassay, high-pressure liquid chromatography, fluorescence polarization immunoassay (FPIA [TDX; Abbott Laboratories, North Chicago, Ill.]), radioimmunoassay (RIA), and fluorescence immunoassay. Based on an analysis of seven standards and of 106 patient samples, all five methods were accurate, and four (bioassay, high-pressure liquid chromatography, FPIA, and RIA) were also precise. The FPIA was the most precise and the fluorescence immunoassay was the least precise of the methods tested; intrarun coefficients of variation for these two methods were 0.9 to 3.0% versus 8.9 to 14.5%, and interrun coefficients of variation were 2.8 to 8.1% versus 12.2 to 16.2%, respectively. The RIA was inconvenient because it required an extra dilution of the specimen being tested and an additional (64 micrograms/ml) vancomycin standard for specimens with 32 to 64 micrograms of vancomycin per ml. Based on its rapid turnaround time and the stability of its standard curve, we believe that the FPIA is the best method currently available to quantitate vancomycin in the clinical laboratory.
机译:我们比较了用于测量血清中万古霉素浓度的五种方法的准确性和准确性:生物测定,高压液相色谱,荧光偏振免疫测定(FPIA [TDX;阿伯特实验室,伊利诺伊州北芝加哥]),放射免疫测定(RIA) ,以及荧光免疫测定。基于对七个标准品和106个患者样品的分析,所有五种方法都是准确的,而四种(生物测定,高压液相色谱,FPIA和RIA)也是精确的。 FPIA是最精确的方法,而荧光免疫测定法是最不精确的方法。这两种方法的内部运行变异系数分别为0.9至3.0%和8.9至14.5%,内部运行变异系数分别为2.8至8.1%与12.2至16.2%。 RIA不方便,因为它需要对待测样品进行额外稀释,并为每毫升含32至64微克万古霉素的样品提供额外的(64微克/毫升)万古霉素标准品。基于其快速的周转时间和标准曲线的稳定性,我们认为FPIA是目前可用于临床实验室中对万古霉素进行定量的最佳方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号