PurposeThe report presents an extraordinary synthesis of customer acceptance procedures (CAP), quality assurance tests (QA) in the treatment of cervical cancer patients, using the first Co-60 Multisource Unit® in Bangladesh. The QA and commissioning required measurements and emergency tests verifying the functional limits of parameters acceptable for the new HDR afterloader. Acceptable limits were: 1) the deviation between specified and measured source strength: ± 3%; 2) the positional accuracy and uniformity: ± 1 mm; 3) the temporal accuracy (i.e. timer error and linearity and end error): ± 1% or 30 sec.; 4) treatment planning system (digitizer and localization software): ± 3% or 1 mm; 5) the distance from line to first dwell position and all the others: 5 mm and 10 mm (± 1 mm).
机译:目的本报告介绍了使用孟加拉国第一台Co-60 Multisource Unit ®治疗宫颈癌患者的客户接受程序(CAP)和质量保证测试(QA)的非凡综合。质量保证和调试需要进行测量和紧急测试,以验证新HDR后装机可接受的参数功能限制。可接受的限值为:1)规定的和测得的光源强度之间的偏差:±3%; 2)位置精度和均匀度:±1 mm; 3)时间精度(即计时器误差,线性度和结束误差):±1%或30秒; 4)治疗计划系统(数字化仪和定位软件):±3%或1 mm; 5)从线到第一停顿位置的距离,以及其他所有距离:5 mm和10 mm(±1 mm)。