A newly developed Digital Radiography (DR) detector has smaller pixel size and higher fill factor than earlier detector models. These technical advantages should theoretically lead to higher sensitivity and higher spatial resolution, thus making dose reduction possible without scarifying image quality compared to previous DR detector versions. To examine whether the newly developed Canon CXDI-70C DR detector provides an improved image quality and/or allows for dose reductions in hand and pelvic bone examinations as well as premature chest examinations, compared to the previous (CXDI-55C) DR detector version. A total of 450 images of a technical Contrast-Detail phantom were imaged on a DR system employing various kVp and mAs settings, providing an objective image quality assessment. In addition, 450 images of anthropomorphic phantoms were taken and analyzed by three specialized radiologists using Visual Grading Analysis (VGA). The results from the technical phantom studies showed that the image quality expressed as IQFINV values was on average approximately 45 % higher with the CXDI-70C detector compared to the CXDI-55C detector. Consistently, the VGA results from the anatomical phantom studies indicated that by using the CXDI-70C detector, diagnostic image quality could be maintained at a dose reduction of in average 30 %, depending on anatomy and kVp level. This indicates that the CXDI-70C detector is significantly more sensitive than the previous model, and supports a better clinical image quality. By using the newly developed DR detector a significant dose reduction is possible while maintaining image quality.
机译:新开发的数字射线照相(DR)检测器比早期的检测器模型具有更小的像素尺寸和更高的填充系数。从理论上讲,这些技术优势应导致更高的灵敏度和更高的空间分辨率,从而与以前的DR检测器版本相比,可以在不降低图像质量的情况下降低剂量。与以前的(CXDI-55C)DR检测器版本相比,要检查新开发的Canon CXDI-70C DR检测器是否提供改进的图像质量和/或允许减少手部和骨盆骨检查以及过早的胸部检查的剂量。在采用各种kVp和mAs设置的DR系统上,共拍摄了450张技术对比度细节模型的图像,从而提供了客观的图像质量评估。此外,由三名专业放射科医生使用视觉评分分析(VGA)拍摄了450张拟人化体模图像并进行了分析。来自技术幻像研究的结果表明,与CXDI-55C检测器相比,CXDI-70C检测器平均以IQFINV值表示的图像质量平均高出约45%。一致地,来自解剖体模研究的VGA结果表明,通过使用CXDI-70C检测器,诊断图像质量可以保持平均降低30%的剂量,具体取决于解剖结构和kVp水平。这表明CXDI-70C检测器比以前的型号灵敏得多,并且支持更好的临床图像质量。通过使用新开发的DR检测器,可以在保持图像质量的同时大幅降低剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号