首页> 美国卫生研究院文献>other >Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia
【2h】

Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia

机译:远端半径和胫骨HR-PQCT图像中对象运动的定量表征

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Image quality degradation due to subject motion is a common artifact affecting in vivo high-resolution peripheral quantitative computed tomography (HR-pQCT) of bone. These artifacts confound the accuracy and reproducibility of bone density, geometry, and cortical and trabecular structure measurements. Observer-based systems for grading image quality and criteria for deciding when to repeat an acquisition and post hoc data quality control remain highly subjective and non-standardized. This study proposed an objective, quantitative technique for measuring subject motion in HR-pQCT acquisitions from the raw projection data, using image similarity measures applied to the parallelized projections at 0° and 180°.A total of 88 HR-pQCT exams with repeated acquisitions of the distal radius (N=54) or distal tibia (N=34) of 49 women (age = 59 >± 14 yr) and 3 men (46 >± 2 yr) were retrospectively evaluated. All images were graded from 1 (no visible motion artifacts) to 5 (severe motion artifacts) according to the manufacturer-suggested image quality grading system. In addition, to serve as the reference case without motion artifacts, two cadaveric wrists and two ankles specimens were imaged twice with repositioning. The motion-induced error was calculated as the percent difference in each bone parameter for the paired scans with and without visually apparent motion artifacts. Quantitative motion estimates (QMEs) for each motion-degraded scan were calculated using two different image similarity measures: sum of squared differences (SSD) and normalized cross-correlation (NCC).The mean values of QMESSD and QMENCC increased with the image quality grade for both radius and tibia. The quality grades were differentiated between grade 2 and 3 using QMESSD, but not with QMENCC, in addition to between grade 4 and 5. Both QMEs correlated significantly to the motion-induced errors in the measurements and their empirical relationship was derived. Subject motion had greater impact on the precision of trabecular structure indices than on the densitometric indices.The results of this study may provide a basis for establishing a threshold for motion artifacts in accordance to the study design as well as a standardized quality control protocol across operators and imaging centers.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号