Maintaining Quality Control Using a Radiological Digital X-ray Dashboard

机译:使用放射数字X射线仪表板维护质量控制

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Repeats are indicators for the quality-imaging manager to schedule additional training and to be used as a basis for dialog with the reading radiologists to improve the service and quality to patients and referring physicians. Through the thoughtful application of software and networking, dose management, X-ray usage, and repeat analysis data can be made available centrally. This provides clinically useful technologist-centric results greatly benefiting an enterprise. This study tracked a radiology department’s use of a digital X-ray dashboard software application. It was discovered that 80% of the exams were performed by only 21% of the technologists and that the technologist with the highest throughput had a personal repeat rate of 6.5% compared to the department average of 7.6%. This study indicated that useful information could be derived and used as a basis for improving the radiology department’s operations and in maintaining high quality standards.
机译:重复是质量成像经理安排其他培训的指标,并用作与放射线放射医生对话的基础,以改善对患者和转诊医生的服务和质量。通过软件和网络的周到应用,可以集中提供剂量管理,X射线使用和重复分析数据。这提供了以临床技术人员为中心的临床有用结果,从而使企业受益匪浅。这项研究追踪了放射科使用数字X射线仪表板软件应用程序的情况。发现80%的考试仅由21%的技术人员执行,并且吞吐量最高的技术人员的个人重复率达到6.5%,而部门平均水平为7.6%。这项研究表明,可以获取有用的信息,并将其用作改善放射科操作和维持高质量标准的基础。