Breast imaging using an amorphous silicon-based full-field digital mammographic system: Stability of a clinical prototype

机译:使用基于非晶硅的全场数字乳腺X线摄影系统进行乳房成像:临床原型的稳定性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An amorphous silicon-based full-breast imager for digital mammography was evaluated for detector stability over a period of 1 year. This imager uses a structured Csl:Tl scintillator coupled to an amorphous silicon layer with a 100-micron pixel pitch and read out by special purpose electronics. The stability of the system was characterized using the following quantifiable metrics: conversion factor (mean number of electrons generated per incident x-ray), presampling modulation transfer function (MTF), detector linearity and sensitivity, defector signal-to-noise ratio (SNR), and American College of Radiology (ACR) accreditation phantom scores. Qualitative metrics such as flat field uniformity, geometric distortion, and Society of Motion Picture and Television Engineers (SMPTE) test pattern image quality were also used to study the stability of the system. Observations made over this 1-year period indicated that the maximum variation from the average of the measurements were less than 0.5% for conversion factor, 3% for presampling MTF over all spatial frequencies, 5% for signal response, linearity and sensitivity, 12% for SNR over seven locations for all 3 target-filter combinations, and 0% for ACR accreditation phantom scores. ACR mammographic accreditation phantom images indicated the ability to resolve 5 fibers, 4 speck groups, and 5 masses at a mean glandular dose of 1.23 mGy. The SMPTE pattern image quality test for the display monitors used for image viewing indicated ability to discern all contrast steps and ability to distinguish line-pair images at the center and corners of the image. No bleeding effects were observed in the image. Flat field uniformity for all 3 target-filter combinations displayed no artifacts such as gridlines, bad detector rows or columns, horizontal or vertical streaks, or bad pixels. Wire mesh screen images indicated uniform resolution and no geometric distortion.
机译:评估了用于数字化乳腺成像的基于非晶硅的全乳腺成像仪在1年内的探测器稳定性。该成像仪使用结构化的Csl:T1闪烁体,该闪烁体以100微米像素间距与非晶硅层耦合,并通过专用电子设备读出。使用以下可量化的指标来表征系统的稳定性:转换因子(每次入射X射线产生的电子平均数),预采样调制传递函数(MTF),检测器线性度和灵敏度,缺陷或信噪比(SNR) )和美国放射学院(ACR)的认证体模分数。定性指标(例如平面场均匀性,几何失真和美国电影电视工程师协会(SMPTE)测试图案图像质量)也用于研究系统的稳定性。在这1年的时间里观察到的结果表明,转换因子与平均测量值的最大差异小于0.5%,对于所有空间频率上的MTF预采样而言,小于3%,对于信号响应,线性和灵敏度,小于5%,小于12%对于所有3个目标滤波器组合,在七个位置上的SNR为0,对于ACR认证体模得分为0%。 ACR乳腺X线照片鉴定模型图像表明,在平均腺体剂量为1.23 mGy的情况下,能够分辨5个纤维,4个斑点组和5个质量。用于图像查看的显示监视器的SMPTE图案图像质量测试表明,能够识别所有对比度步骤,并能够区分图像中心和角落的线对图像。在图像中未观察到出血效果。所有3个目标滤波器组合的平面场均匀性均未显示任何伪影,例如网格线,检测器行或列不良,水平或垂直条纹或不良像素。丝网屏幕图像显示均匀的分辨率,没有几何变形。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号